Radiologist assessment of breast density by BI-RADS categories versus fully automated volumetric assessment

机译:放射科医生通过BI-RADS类别对乳房密度的评估与全自动体积评估

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

OBJECTIVE. The objective of our study was to estimate mammographic breast density using a fully automated volumetric breast density measurement method in comparison with BI-RADS breast density categories determined by radiologists. MATERIALS AND METHODS. A total of 791 full-field digital mammography examinations with standard views were evaluated by three blinded radiologists as BI-RADS density categories 1-4. For fully automated volumetric analysis, volumetric breast density was calculated with fully automated software. The volume of fibroglandular tissue, the volume of the breast, and the volumetric percentage density were provided. RESULTS. The weighted overall kappa was 0.48 (moderate agreement) for the three radiologists' estimates of BI-RADS density. Pairwise comparisons of the radiologists' measurements of BI-RADS density revealed moderate to substantial agreement, with kappa values ranging from 0.51 to 0.64. There was a significant difference in mean volumetric breast density among the BI-RADS density categories, and the mean volumetric breast density increased as the BI-RADS density category increased (p < 0.001). A significant positive correlation was found between BI-RADS categories and fully automated volumetric breast density (? = 0.765, p < 0.001). CONCLUSION. Our study showed good correlation of the fully automated volumetric method with radiologist-assigned BI-RADS density categories. Mammographic density assessment with the fully automated volumetric method may be used to assign BI-RADS density categories.
机译:目的。我们研究的目的是与由放射科医生确定的BI-RADS乳房密度类别相比,使用全自动容积乳房密度测量方法来估计乳房X线照片的乳房密度。材料和方法。由三位不知情的放射科医生按照BI-RADS密度类别1-4评估了总共791次标准视野的全乳腺钼靶X光检查。对于全自动体积分析,使用全自动软件计算乳房体积。提供了纤维腺组织的体积,乳房的体积和体积百分比密度。结果。三位放射线医师对BI-RADS密度的估计,加权总kappa为0.48(中度一致)。放射科医生对BI-RADS密度的测量结果的成对比较表明,中度至基本一致,kappa值介于0.51至0.64之间。 BI-RADS密度类别之间的平均乳房体积密度存在显着差异,并且随着BI-RADS密度类别的增加,乳房平均体积密度也随之增加(p <0.001)。在BI-RADS类别和全自动容积乳腺密度之间发现显着正相关(?= 0.765,p <0.001)。结论。我们的研究表明,全自动体积法与放射科医生指定的BI-RADS密度类别具有良好的相关性。具有全自动体积法的乳房X线密度评估可用于分配BI-RADS密度类别。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号