In Vitro Examination of the VentriFlo True Pulse Pump for Failing Fontan Support

机译:对Ventriflo真正脉冲泵的体外检查失败Fontan支持

获取原文
获取原文并翻译 | 示例

摘要

The current methodology of Fontan palliation results in a one "pump" circulatory system with passive flow to the lungs. Inherent hemodynamic differences exist between a biventricular circulatory system and this modified physiology, leading to a host of long-term complications. Mechanical circulatory support (MCS) is a potential option to combat these pathophysiological conditions. In this study, we examine the VentriFlo True Pulse Pump as a MCS option to support a failing Fontan patient. An in vitro circulatory loop was used to model a failing Fontan patient, reproducing pathophysiological pressures and flow rates. The VentriFlo True Pulse Pump was positioned as a right sided support, testing multiple cannulation and baffle restriction strategies, as well as various pumping parameters including flow rate, frequency, stroke volume and the ejection to filling time ratio. A 10 mm Hg decrease in IVC pressure and 0.75 L/min increase in cardiac output were achieved using a complete baffle restriction strategy. Additional cannulation and banding strategies were not as successful. Pump flow rate and frequency significantly impacted hemodynamics, while the ejection to filling time ratio did not. Though not ideal, complete baffle restriction was necessary to achieve successful support. The ability to tune individual pumping parameters for a given MCS device will have a substantial impact on the pressures and flow augmentation seen in a Fontan circulation. Both future pump design and off-label VADs for Fontan use should consider the pump configuration and parameter combinations presented here, which offered successful support.
机译:Fontan Palliation的目前的方法是一个“泵”循环系统,具有肺部被动流动。生物循环系统和这种修饰的生理学之间存在固有的血流动力学差异,导致了一系列长期并发症。机械循环支持(MCS)是对抗这些病理生理病症的潜在选择。在这项研究中,我们将Neverriflo真脉冲泵作为MCS选项,以支持失败的Fontan患者。体外循环回路用于模拟失败的Fontan患者,再现病理生理学压力和流速。 Ventriflo真脉冲泵定位为右侧支撑,测试多个插管和挡板限制策略,以及各种泵送参数,包括流速,频率,行程体积和喷射到填充时间比。使用完整的挡板限制策略,实现了10mm的IVC压力和0.75L / min的0.75升/分钟的增加。额外的插管和绑带策略并不成功。泵流量和频率显着撞击血流动力学,而弹出到填充时间比例没有。虽然不是理想的,但完全挡板限制是实现成功支持所必需的。对给定MCS装置的各个泵送参数调谐的能力将对在Fontan循环中看到的压力和流量增强具有显着影响。 Fontan使用的未来泵设计和非标签VAD都应考虑此处提供的泵配置和参数组合,从而提供成功的支持。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号