首页> 外文期刊>Acta tropica: Journal of Biomedical Sciences >Evaluation of the LH780 hematology analyzer for detection and therapeutic monitoring of malaria: Cross-reactivity with nucleated RBCs
【24h】

Evaluation of the LH780 hematology analyzer for detection and therapeutic monitoring of malaria: Cross-reactivity with nucleated RBCs

机译:LH780血液分析仪用于疟疾检测和治疗监测的评估:与有核红细胞的交叉反应

获取原文
获取原文并翻译 | 示例

摘要

We evaluated the diagnostic usefulness of the LH780 Coulter blood cell counter for detection and therapeutic monitoring of malaria including cross-reactivity with nucleated RBC (NRBC) samples. A total of 405 patients (43 patients with Plasmodium vivax malaria and the control group of 120 healthy subjects, 111 patients with fever, and 131 patients with NRBCs) were analyzed with routine CBC using the LH780. We analyzed the CBC results according to three selected parameters: an abnormal peak in the WBC histogram before 35 fL, the presence of red dots in the nonwhite cell zone of 2D WBC Diff Data plot, and platelet-related flags suggesting platelet clumps or giant platelets. Of the 43 malaria samples collected at diagnosis, an abnormal peak (>= 2.2 mm) was present in 93.0% (95% confidential interval (CI), 80.9-98.5%). Of all samples, 97.7% (95% CI, 87.7-99.9%) exhibited red dots, and platelet-related flags were observed in 81.4% (95% CI, 66.6-91.6%). The specificity of these three selected parameters was 83.1% (95% CI, 78.9-86.9%), 77.3% (95% CI, 72.7-81.6%), and 90.1% (95% CI, 86.5-92.9%), respectively. The abnormal peak (>= 2.2 mm) showed moderate correlation with parasite level (r= 0.79). The three selected LH780 parameters were useful for identifying malaria in healthy subjects and febrile patients, but unsatisfactory for discriminating malaria in NRBC samples. The parameters showed a substantial proportion of false positives in the NRBC group, ranging from 26.7% to 49.6%. Therefore, microscopic confirmation will be necessary for application of these parameters for malaria screening and treatment monitoring. (C) 2016 Elsevier B.V. All rights reserved.
机译:我们评估了LH780 Coulter血细胞计数器对疟疾的检测和治疗监测(包括与有核RBC(NRBC)样品的交叉反应性)的诊断实用性。使用LH780,通过常规CBC对405例患者(43例间质疟原虫疟疾患者和120名健康受试者的对照组,111例发烧患者和131例NRBC患者)进行了常规CBC分析。我们根据三个选定的参数分析了CBC结果:35 fL之前的WBC直方图中的异常峰,二维WBC Diff数据图的非白细胞区域中存在红点,以及与血小板相关的标志提示血小板聚集或巨血小板。在诊断时收集的43个疟疾样本中,有93.0%(95%可信区间(CI),80.9-98.5%)出现异常峰值(> = 2.2 mm)。在所有样品中,有97.7%(95%CI,87.7-99.9%)出现红点,并且在81.4%(95%CI,66.6-91.6%)中观察到血小板相关的标志。这三个选定参数的特异性分别为83.1%(95%CI,78.9-86.9%),77.3%(95%CI,72.7-81.6%)和90.1%(95%CI,86.5-92.9%)。异常峰(> = 2.2 mm)显示与寄生虫水平适度相关(r = 0.79)。选择的三个LH780参数可用于识别健康受试者和高热患者中的疟疾,但不能令人满意地区分NRBC样本中的疟疾。参数显示NRBC组中假阳性的比例很高,范围从26.7%至49.6%。因此,对于将这些参数应用于疟疾筛查和治疗监测,必须进行显微镜确认。 (C)2016 Elsevier B.V.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号