Accuracy of PurePulse photoplethysmography technology of Fitbit Charge 2 for assessment of heart rate during sleep

机译:FITBET CHARGET的Purepulse PhotoPretysMomography技术的准确性,用于评估睡眠期间心率

获取原文
获取原文并翻译 | 示例
       

摘要

Elevated asleep heart rate (HR) is a risk factor for cardiovascular disease and other-cause morbidity and mortality. We assessed the accuracy of Fitbit Inc. PurePulse (R) photoplethysmography with reference to three-lead electrocardiography (ECG) in determining HR during sleep. HR of 35 (17 female) healthy adults 25.1 +/- 10.6 years of age (mean +/- SD) was continuously recorded throughout a single night of sleep. There was no significant difference in asleep HR mean (0.09 beats per minute [bpm], P = 0.426) between Fitbit photoplethysmography and ECG; plus, there was excellent intraclass correlation (0.998) and narrow Bland-Altman agreement range (2.67 bpm). The regression analysis of Bland-Altman plot of mean asleep HR indicates Fitbit tends to slightly overestimate reference values in the lower range of HR (HR < 50 bpm) by 0.51 bpm and slightly underestimate reference values in the higher range of HR (HR > 80 bpm) by 0.63 bpm. Mixed model analysis of epoch-by-epoch (5-min epochs) asleep HR showed significant "U" shape trend (P < 0.001) in amount of Fitbit error (absolute amount of difference between ECG and Fitbit values regardless of overestimation or underestimation) in regard to HR, i.e. smaller error in the medium range of HR (60-80 bpm) and slightly larger error for lower (<60 bpm) and higher (>80 bpm) ranges of HR. However, effect of age, body mass index, gender, and subjective sleep quality measured by Pittsburgh sleep quality index (good/poor sleepers) on error in estimating HR by the Fitbit method was not significant. It is concluded that Fitbit photoplethysmography suitably tracks HR during sleep in healthy young adults.
机译:睡眠心率升高(HR)是心血管疾病和其他引起的发病率和死亡率的危险因素。我们评估了Fitbit Inc.Putbulse(R)光增性肌拍的准确性,参考睡眠期间的三铅心电图(ECG)。 35(17名女性)健康成人的人力资源25.1 +/- 10.6岁(平均+/- SD)在一晚睡眠中不断录制。 Fitbit PhotoprysmographysMocography和ECG之间的睡眠HR平均值(0.09节拍,P = 0.426)没有显着差异;此外,有优异的脑内相关性(0.998)和窄的Bland-Altman协议范围(2.67 BPM)。平均睡眠HR的Bland-Altman图的回归分析表明FitBit趋于稍微高估在低于HR(HR <50bpm)的参考值0.51bpm,略微低估HR的参考值(HR> 80 BPM)0.63 bpm。 EPOCH-BY-EPOCH(5分钟时期)的混合模型分析睡眠HR在Fitbit误差量的数量显示出显着的“U”形趋势(P <0.001)(ECG和Fitbit值之间的绝对差异,无论高估还是低估)关于HR,即HR的中等范围内的误差(60-80bpm)和低(<60bpm)的稍微误差略大误差,高于(> 80bpm)的HR。然而,匹兹堡睡眠质量指数(良好/可怜的睡眠者)通过Fitbit方法估算HR误差衡量的年龄,体重指数,性别和主观睡眠质量的影响并不重要。它的结论是,Fitbit PhotoPrishysmography适当地在健康的年轻成年人睡眠期间适当地跟踪HR。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号