Calibration by commutable control materials is able to reduce inter-method differences of current high-performance methods for HbA 2

机译:可通勤控制材料校准能够降低HBA 2的当前高性能方法的方法间差异

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Abstract Background Most of the current methods used for the determination of HbA 2 seem not well aligned. A comparison among the best performing techniques and the commutability of some control materials currently available and under development has been evaluated. Methods Forty blood samples were analyzed in duplicate over two separate days by different HPLC and capillary electrophoresis systems. The commutabilities of different control materials (NIBSC WHO reagent, Bio-Rad Lyphochek, and home prepared lyophilized controls RP1–3) have been assessed by analyzing the controls in quadruplicate over two consecutive days together with the blood samples. Results The mean within-run imprecision of HbA 2 measurement on blood samples (CV, %) was between 0.6% and 10.1% for HbA 2 values 2 ≥3.5%. The different methods were highly correlated (r between 0.9941 and 0.9995) although biased each other. The NIBSC WHO reagent was found not commutable in 15 over 28 comparisons, the Lyphochek 2 in 18/28, and RP3 in 4/28. Recalibration of all methods by RP1 and RP2 materials was able to reduce the overall variability from 6.8% to 3.4% at HbA 2 ≤3.0% and from 6.7% to 3.0% at HbA 2 ≥4.6%. Conclusion The use of adequate commutable control materials as calibrators may reduce the inter-method variability of routine methods to an extent closer to the current analytical goals of bias based on biological variability. Highlights ? Current high-performance methods for HbA2 are well correlated each other although biased. ? Most methods meet the minimum goals for imprecision based on biological variability. ? The only WHO reference reagent for HbA2 is not commutable for the majority of methods. ? Home-prepared lyophilized control materials are commutable with the vast majority of routine methods.
机译:摘要背景用于测定HBA 2的最新方法看起来不太顺序。已经评估了当前可用的一些控制材料的最佳表现技术和可在开发的可扩展性的比较。方法通过不同的HPLC和毛细管电泳系统在两个单独的一天中重复分析四十血样。通过分析与血样连续两个天中的双包装中的对照进行分析,已经评估了不同控制材料(NIBSC的铌试剂,生物-Rad Lyphochek和家庭制备的冻干对照RP1-3)。结果HBA 2对血液样品(CV,%)测量的平均局部不测性为HBA 2值2≥3.5%的0.6%和10.1%。不同的方法高度相关(r为0.9941和0.9995),尽管彼此偏置。在18/28中的兰白克28例比比较超过28次比较中,发现试剂的NIBSC不可通货,并在4/28中的RP3。 RP1和RP2材料的所有方法的重新校准能够将总变异从HBA2≤3.0%的6.8%降低至3.4%,HBA 2≥4.6%的6.7%至3.0%。结论使用适当的可通道控制材料作为校准器可以降低常规方法的互相互变性,以基于生物可变性更接近当前偏置的分析目标。强调 ?虽然偏置,但HBA2的电流高性能方法彼此良好。还大多数方法根据生物可变性符合不精确的最低目标。还HBA2的唯一参考试剂对大多数方法都不可通货。还家庭准备的冻干控制材料与绝大多数常规方法通货。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号