Concordance between laboratories in metal ion testing in patients with metal-on-metal hip implants

机译:金属髋桩植入物患者金属离子试验中的实验室之间的一致性

获取原文
获取原文并翻译 | 示例
           

摘要

Background Testing of whole blood or serum metal ion levels has become an important part of assessing and monitoring the performance of metal-on-metal bearings, both in hip resurfacing arthroplasty and in total hip replacement. The aim of this study was to determine the concordance between 2 laboratories testing cobalt and chromium ion levels in patients with metal-on-metal bearings. Methods Serum and whole blood samples from patients who had undergone metal-on-metal resurfacing or large-diameter total hip arthroplasty were tested for cobalt and chromium ions in laboratory A (a recognized laboratory) and laboratory B (tasked with testing clinical specimens). Laboratory A performed cobalt and chromium testing on whole blood, and laboratory B performed cobalt testing on whole blood and chromium testing on serum. Results Samples from 104 patients were tested. Laboratory B reported lower whole blood cobalt levels than laboratory A. Furthermore, laboratory A reported that all patients had elevated whole blood cobalt ion levels compared to the normal reference values for the laboratory, whereas laboratory B reported that 46 patients (44.2%) had whole blood cobalt ion levels within the normal reference range for the laboratory. Conclusion This comparative study highlights the importance of using a single laboratory for metal ion testing, as values generated from different laboratories may not be directly comparable. With recent literature suggesting that whole blood cobalt levels as low as 1 ppb may be a predictor of adverse reactions to metal debris, accurate clinical measurement needs to be increasingly exact.
机译:背景技术全血或血清金属离子水平的测试已成为评估和监测金属轴承的性能的重要组成部分,均在髋关节复制关节置换术和总髋关节替代品中进行金属轴承的性能。本研究的目的是确定金属金属轴承患者的2实验室测试钴和铬离子水平之间的一致性。方法在实验室A(公认的实验室)和实验室B(任务用测试临床标本)中,在经过金属金属覆盖物或大直径总髋关节置换术中进行血清和大直径总髋关节置换术的血清和全血液样本。实验室A对全血的钴和铬检测,实验室B对全血和血清铬检测进行了钴测试。结果来自104名患者的样品。实验室B报告了比实验室A的全血钴水平降低了。此外,实验室A据报道,与实验室的正常参考值相比,所有患者均升高了全血钴离子水平,而实验室B报告了46名患者(44.2%)整体实验室的正常参考范围内的血液钴离子水平。结论该比较研究强调了使用单一实验室对金属离子测试的重要性,因为从不同实验室产生的值可能没有直接比较。随着最近的文献表明,整个血液钴水平低至1 ppb可能是对金属碎片不良反应的预测因子,准确的临床测量需要越来越精确。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号