Comparison of the Thrombogenicity of a Bare and Antithrombogenic Coated Flow Diverter in an In Vitro Flow Model

机译:裸露涂层涂层流动分流器在体外流动模型中的血栓形成性比较

获取原文
获取原文并翻译 | 示例
           

摘要

Background Dual antiplatelet therapy is a pre-requisite for flow diverter (FD) implantation. The purpose of this study was to assess the thrombogenicity of the p48 FD, coated with the newly developed phenox Hydrophilic Polymer Coating (p48_HPC, phenox GmbH, Germany) in comparison with uncoated p48 FDs in an in vitro flow model (Chandler Loop). Methods p48 and p48_HPC FDs were implanted into silicon tubes filled with whole human blood and incubated at 37 degrees C under pulsating flow. After 120 min, platelet count was determined in the blood. Platelet activation markers (PAR1) and formation of microparticles were analyzed in a flow cytometer. Fluorescence microscopy of CD42a positive cells and scanning electron microscopy was used to detect adherent platelets on the wire surface. Results Platelets in contact with the uncoated p48 FDs are significantly more activated than those incubated with p48_HPC (73 +/- 9% vs. 65 +/- 6%, p < 0.05) and release more microparticles (1.8 +/- 0.5 vs. 1.4 +/- 0.4, p < 0.05). The platelet count after 120-min circulation in the Chandler Loop was significantly lower for the uncoated p48 compared to the p48_HPC indicating significantly greater adherence of the platelets to the p48 (71 +/- 8% vs. 87 +/- 5%, p < 0.05). SEM and fluorescent antibody imaging revealed minimal platelet adherence to the surface of the p48_HPC compared to the uncoated p48. Conclusion The pHPC coating significantly reduces thrombogenicity of the p48 FD. This may help to reduce the risk of thromboembolic complications when using these devices. A reduction in antiplatelet therapy may be possible.
机译:背景技术双抗血小板疗法是流动转向器(FD)植入的预先要求。本研究的目的是评估P48 FD的血栓形成性,涂有新开发的苯氧基亲水性聚合物涂层(P48_HPC,德国),与在体外流量模型(Chandler Loop)中的未涂覆的P48 FD相比。方法将P48和P48_HPC FDS植入填充有全部人血液的硅管中,并在脉动流动下在37摄氏度下孵育。 120分钟后,血液中测定血小板计数。在流式细胞仪中分析血小板活化标记物(PAR1)和微粒的形成。 CD42A阳性细胞和扫描电子显微镜的荧光显微镜用于检测线表面上的粘附血小板。结果与未涂覆的P48 FDS接触的血小板显着更激活,而不是与p48_HPC孵育的血小板(73 +/- 9%,P <0.05)并释放更多的微粒(1.8 +/- 0.5 Vs. 1.4 +/- 0.4,p <0.05)。与P48_HPC相比,未涂覆的P48,在Chandler回路中120分钟循环后的血小板计数显着降低了P48_HPC对P48的显着提高血小板粘附(71 +/- 8%,P. +/- 5%,P <0.05)。与未涂覆的P48相比,SEM和荧光抗体成像显示出对P48_HPC表面的最小血小板粘附。结论PHPC涂层显着降低了P48 FD的血栓形成性。这可能有助于在使用这些设备时降低血栓栓塞并发症的风险。抗血小板治疗的降低可能是可能的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号