首页> 外文会议>Society for Biomaterials annual meeting and exposition >Thrombogenicity Evaluation of Medical Device Materials with an In Vitro Flow Loop System Under Varying Flow Conditions
【24h】

Thrombogenicity Evaluation of Medical Device Materials with an In Vitro Flow Loop System Under Varying Flow Conditions

机译:可变流量条件下体外循环回路系统评估医疗器械材料的血栓形成性

获取原文

摘要

The data suggest that flow rate substantially impacts the thrombogenicity differentiation of biomaterials. For the current system, 200 mL/min provided the best distinction between the materials. The variability in thrombosis for in vitro blood flow loops may be decreased by using controlled donor-specific heparin concentrations and system-specific flow rates. The higher hemolysis levels produced at 300 mL/min may contribute to non-material mediated coagulation and therefore a flow rate at or above this level may not be appropriate for examining device thrombogenicity in this system.
机译:数据表明流速显着影响生物材料的血栓形成分化。对于当前系统,200 mL / min提供了两种材料之间的最佳区别。通过使用受控的供体特异性肝素浓度和系统特异性流速,可以减少体外血流循环的血栓形成变异性。以300 mL / min产生的较高溶血水平可能会导致非物质介导的凝结,因此,此水平或更高的流速可能不适用于检查该系统中的装置血栓形成性。