Comparison between coronal reformatted images and direct coronal CT images of the swine lung specimen: assessment of image quality with 64-detector row CT.

机译:猪肺试样的冠状重新格式化图像与直冠CT图像的比较:64探测仪行CT的图像质量评估。

获取原文
获取原文并翻译 | 示例
           

摘要

The aims of this study were to compare the image quality of coronal multiplanar reconstruction (MPR) images from axial spiral images with that of direct coronal spiral and sequential images, and to estimate and analyse the effect of an incremental change on the image quality using 64-detector row CT. 12 swine lungs were used. Five kinds of images from each lung specimen were obtained using 64-detector row CT. All images were analysed by categories and grades, and the direct coronal sequential images were used as the reference standard for the image quality. Statistical analysis was performed for the following categories: (i) inter-observer reliability, (ii) interaction between the observers and images, (iii) image analysis, (iv) anatomical structural analysis of each observer, (v) stair-step artefact and (vi) background noise. The overall image quality and the image quality of all anatomical structures of coronal MPR images with 0.67 mm slice increments were inferior to the image quality of the other images; this difference was statistically significant (p<0.05). Stair-step artefact was detected on coronal MPR images, and was more prominent on coronal MPR images with 0.67 mm slice increments than on coronal MPR images with 0.34 mm slice increments. The most severe background noise was detected on the direct coronal sequential images, but there was no significant difference between the direct coronal sequential images and the direct coronal spiral images. Background noise was least prominent on coronal MPR images with 0.67 mm slice increments. The increment process is important for improving the image quality of MPR images even when using 64-detector row CT. Coronal MPR images with 0.34 mm slice increments using 64-detector row CT showed a similar image quality to that obtained from the direct coronal images, and can be used instead. This means that the coronal MPR images obtained with 64-detector row CT could be as useful for evaluating the lung parenchyma as the axial high-resolution CT images.
机译:本研究的目的是将来自轴向螺旋图像的冠状多平坦重建(MPR)图像的图像质量与直接冠状螺旋和顺序图像的图像进行比较,并估计和分析使用64的增量变化对图像质量的影响-dector行ct。使用12种猪肺。使用64检测器排CT获得来自每个肺标本的五种图像。通过类别和等级分析所有图像,直接冠状序列图像被用作图像质量的参考标准。对以下类别进行统计分析:(i)观察者间可靠性(ii)观察者和图像之间的相互作用,(iii)图像分析(iv)每个观察者的解剖结构分析,(v)阶梯阶段(vi)背景噪音。冠状MPR图像的所有解剖结构的整体图像质量和图像质量与0.67mm的切片增量差不多到其他图像的图像质量;这种差异有统计学意义(P <0.05)。在冠状MPR图像上检测到阶梯人工制品,并且在冠状MPR图像上更加突出,而不是0.67mm的切片增量,而不是0.34mm切片增量的冠状MPR图像。在直接冠状序列图像上检测到最严重的背景噪声,但直接冠状顺序图像和直冠螺旋图像之间没有显着差异。背景噪声在冠状MPR图像上最小突出,切片增量0.67毫米。即使在使用64检测器行CT时,增量过程对于提高MPR图像的图像质量也很重要。使用64探测器行CT的0.34mm切片增量的冠状MPR图像显示了与从直接冠状图像获得的类似图像质量,并且可以代替。这意味着用64检测器行CT获得的冠状MPR图像可以是评估肺实质作为轴向高分辨率CT图像的有用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号