首页> 外文期刊>Journal of Materials Science >Fracture analysis of ceramic femoral head in hip arthroplasty based on microdamage monitoring using acoustic emission
【24h】

Fracture analysis of ceramic femoral head in hip arthroplasty based on microdamage monitoring using acoustic emission

机译:基于声发射微损伤监测的髋关节置换术中股骨头陶瓷的断裂分析

获取原文
获取原文并翻译 | 示例
           

摘要

Damage accumulation during compression testing of the alumina femoral head used in hip arthroplasty was monitored using an acoustic emission (AE) technique. Because a number of mechanical noises due to friction disturbed the AE measurement during the test using a conventional testing configuration standardized in ISO 7206, the testing apparatus was modified so that AE signals from microdamage were detected successfully. During the compression tests of femoral heads, a rapid increase in AE energy was observed before the final fracture. Dye penetration observations demonstrated that the rapid increase in AE energy correlated with the formation of maincrack which leads to the final fracture. The stress distribution in femoral heads was analyzed using finite element analysis (FEA). The critical stress for maincrack formation and the final fracture strength were determined. Finally, fundamental insights into the development of a technique for assessing the long-term reliability of ceramic femoral heads were obtained.
机译:使用声发射(AE)技术监测在髋关节置换术中使用的氧化铝股骨头的压缩测试过程中的损伤累积。由于在使用ISO 7206标准化的常规测试配置进行的测试过程中,由于摩擦而产生的许多机械噪声干扰了AE测量,因此修改了测试设备,以便成功检测到来自微损伤的AE信号。在股骨头加压试验中,在最终骨折之前观察到AE能量迅速增加。染料渗透观察表明,AE能量的快速增加与主裂纹的形成相关,导致最终断裂。使用有限元分析(FEA)分析了股骨头中的应力分布。确定了主裂纹形成的临界应力和最终的断裂强度。最后,获得了对陶瓷股骨头长期可靠性评估技术发展的基本见识。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号