Tear glucose measurements have been suggested as a potential alternative to blood glucose monitoring for diabetic patients. While previous work has reported that there is a correlation between blood and tear glucose levels in humans, this link has not been thoroughly established and additional clinical studies are needed. Herein, we evaluate the potential of using commercial blood glucose test strips to measure glucose in tears. Of several blood glucose strips evaluated, only one brand exhibits the low detection limit required for quantitating glucose in tears. Calibration of these strips in the range of 0-100 μM glucose with an applied potential of 150 mV to the working electrode yields a sensitivity of 0.127 nA/μM and a limit of quantitation (LOQ) of 9 μM. The strips also exhibit ≤13% error (n = 3) for 25, 50, and 75 μM glucose in the presence of 10 μM acetaminophen, 100 μM ascorbic acid, and 100 μM uric acid. Measurements of glucose in tears from nine normal (nondiabetic) fasting human subjects using strips yielded glucose values within the range of 5-148 μM (mean = 47 μM, median = 43 μM), similar to those for human tears reported by others with more complex LC-MS methods. The glucometer strip method could facilitate more clinical studies to determine whether tear glucose and blood glucose levels sufficiently correlate for application to routine measurements in tears to supplement blood glucose testing. This would be especially helpful for children, adolescents, other Type 1 diabetics, and also for Type 2 diabetics who require treatment with insulin and cannot tolerate multiple finger sticks per day.
机译:有人建议对糖尿病患者进行泪糖测量,以代替血糖监测。尽管先前的工作已报道人类血液和泪液葡萄糖水平之间存在相关性,但这种联系尚未得到充分建立,需要进行其他临床研究。本文中,我们评估了使用商业血糖试纸测量眼泪中葡萄糖的潜力。在评估的几条血糖条中,只有一个品牌显示出定量眼泪中葡萄糖所需的低检测限。在0-100μM葡萄糖范围内对这些条进行校准,并向工作电极施加150 mV的电势,灵敏度为0.127 nA /μM,定量限(LOQ)为9μM。在10μM对乙酰氨基酚,100μM抗坏血酸和100μM尿酸存在的情况下,对于25、50和75μM葡萄糖,试纸条还显示出≤13%的误差(n = 3)。使用试纸对九名正常(非糖尿病)禁食的人类受试者的眼泪中的葡萄糖进行测量,得出的葡萄糖值在5-148μM范围内(平均值= 47μM,中位数= 43μM),与其他人报道的针对人类眼泪的葡萄糖值相似复杂的LC-MS方法血糖仪剥离法可以促进更多的临床研究,以确定泪液葡萄糖和血糖水平是否足够相关,可用于常规的泪液测量以补充血糖测试。这对于儿童,青少年,其他1型糖尿病患者以及需要胰岛素治疗且每天不能耐受多根手指的2型糖尿病患者尤其有用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号