首页> 外文期刊>Diabetes technology & therapeutics >Accuracy of subcutaneous continuous glucose monitoring in critically ill adults: Improved sensor performance with enhanced calibrations
【24h】

Accuracy of subcutaneous continuous glucose monitoring in critically ill adults: Improved sensor performance with enhanced calibrations

机译:危重成年人皮下连续血糖监测的准确性:通过增强校准提高传感器性能

获取原文
获取原文并翻译 | 示例
           

摘要

Objective: Accurate real-time continuous glucose measurements may improve glucose control in the critical care unit. We evaluated the accuracy of the FreeStyle? Navigator? (Abbott Diabetes Care, Alameda, CA) subcutaneous continuous glucose monitoring (CGM) device in critically ill adults using two methods of calibration. Subjects and Methods: In a randomized trial, paired CGM and reference glucose (hourly arterial blood glucose [ABG]) were collected over a 48-h period from 24 adults with critical illness (mean±SD age, 60±14 years; mean±SD body mass index, 29.6±9.3 kg/m2; mean±SD Acute Physiology and Chronic Health Evaluation score, 12±4 [range, 6-19]) and hyperglycemia. In 12 subjects, the CGM device was calibrated at variable intervals of 1-6 h using ABG. In the other 12 subjects, the sensor was calibrated according to the manufacturer's instructions (1, 2, 10, and 24 h) using arterial blood and the built-in point-of-care glucometer. Results: In total, 1,060 CGM-ABG pairs were analyzed over the glucose range from 4.3 to 18.8 mmol/L. Using enhanced calibration median (interquartile range) every 169 (122-213) min, the absolute relative deviation was lower (7.0% [3.5, 13.0] vs. 12.8% [6.3, 21.8], P0.001), and the percentage of points in the Clarke error grid Zone A was higher (87.8% vs. 70.2%). Conclusions: Accuracy of the Navigator CGM device during critical illness was comparable to that observed in non-critical care settings. Further significant improvements in accuracy may be obtained by frequent calibrations with ABG measurements.
机译:目的:准确,实时,连续的血糖测量可以改善重症监护病房的血糖控制。我们评估了FreeStyle的准确性吗?航海家? (Abbott Diabetes Care,Alameda,CA)在危重成年人中使用两种校准方法进行皮下连续葡萄糖监测(CGM)设备。受试者和方法:在一项随机试验中,在48小时内从24位重症患者(平均±SD年龄,60±14岁;平均值±)中收集了配对的CGM和参考葡萄糖(每小时动脉血葡萄糖[ABG])。 SD体重指数为29.6±9.3 kg / m2;平均值±SD急性生理和慢性健康评估得分为12±4 [范围,6-19])和高血糖症。在12位受试者中,使用ABG以1-6小时的可变间隔校准了CGM设备。在其他12位受试者中,使用制造商的说明书(1、2、10和24小时)使用动脉血和内置即时血糖仪对传感器进行了校准。结果:在葡萄糖范围为4.3至18.8 mmol / L的条件下,共分析了1,060对CGM-ABG。每169(122-213)分钟使用增强的校准中位数(四分位间距),绝对相对偏差会降低(7.0%[3.5,13.0]与12.8%[6.3,21.8],P <0.001),并且百分比Clarke误差网格A区中的积分更高(87.8%对70.2%)。结论:危重病患者期间Navigator CGM设备的准确性与非危重医疗机构中观察到的准确性相当。通过使用ABG测量进行频繁的校准,可以进一步提高准确性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号