PURPOSE: To evaluate the acute effects of the Amplatz thrombectomy device (ATD) on peripheral venous valves in a canine model. MATERIALS AND METHODS: ATD thrombectomy was performed in 17 veins, and control experiments with use of an 8-F sheath-dilator were performed in four veins. Prethrombectomy ascending venography was performed, followed by device passage across the vein segment. Post-thrombectomy ascending venography was then performed, followed by heparinization and euthanasia. The treated veins were carefully explanted and stored in formaldehyde for histopathologic examination. Severity of valve injury was graded on a scale of 0 to 4. RESULTS: In ATD-treated veins: 10 veins sustained no injury [grade 0] (diameter, 6.7 mm +/- 1.7; antegrade/retrograde approach, 5/5), five veins sustained mild injury [grade 1-2] (diameter, 5.2 mm +/- 0.8; antegrade/retrograde, 3/2), while the remaining two veins sustained moderate-to-severe injury [grade 3-4] (diameter, 5 and 6 mm; antegrade/retrograde, 1/1). In sheath-dilator treated veins: no injury [grade 0] in any of the four treated veins (mean diameter, 5.5 mm +/- 0.6; all retrograde). In ATD-treated veins, valve injury (of any grade) was significantly more frequent in veins 6 mm or less in diameter than in veins at least 7 mm in diameter (seven of 12 vs zero of five; P < .03). There was no significant association between thrombectomy approach and injury grade. CONCLUSION: Veins 7 mm or greater in diameter were associated with no significant valve injury during ATD thrombectomy. However, long-term and short-term effects on valvular function will need to be assessed.
机译:目的:评估犬模型中Amplatz血栓切除术(ATD)对外周静脉瓣膜的急性作用。材料与方法:在17条静脉中进行ATD血栓切除术,并在4条静脉中进行使用8-F鞘膜扩张器的对照实验。进行血栓切除术前静脉造影,然后使器械穿过静脉段。然后进行血栓切除术后上行静脉造影,然后进行肝素化和安乐死。将经治疗的静脉小心移出,并保存在甲醛中以进行组织病理学检查。瓣膜损伤的严重程度分级为0到4。结果:在经ATD处理的静脉中:10条静脉没有受到损伤[0级](直径,6.7毫米+/- 1.7;顺行/逆行方法,5/5)。 ,五条静脉受到轻度损伤[1-2级](直径5.2 mm +/- 0.8;顺行/逆行3/2),其余两条静脉受到中度至重度损伤[3-4级](直径,分别为5和6毫米;顺/逆,1/1)。在鞘膜扩张器治疗的静脉中:四根治疗的静脉中的任何一根均无损伤(0级)(平均直径,5.5 mm +/- 0.6;全部逆行)。在经ATD处理的静脉中,直径不超过6毫米的静脉比任何直径至少7毫米的静脉明显更频繁地发生瓣膜损伤(任何级别)(12个中的7个vs 5个中的零; P <.03)。血栓切除方法与损伤程度之间无显着关联。结论:直径为7mm或更大的静脉与ATD血栓切除术期间无明显瓣膜损伤有关。但是,将需要评估对瓣膜功能的长期和短期影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号