首页> 外文期刊>Journal of vascular and interventional radiology: JVIR >Use of a catheter-based system to measure blood flow in hemodialysis grafts during angioplasty procedures.
【24h】

Use of a catheter-based system to measure blood flow in hemodialysis grafts during angioplasty procedures.

机译:使用基于导管的系统在血管成形术过程中测量血液透析移植物中的血流量。

获取原文
获取原文并翻译 | 示例

摘要

PURPOSE: The goals of this investigation were to evaluate the accuracy and reliability of the Angioflow meter system with use of in vitro and in vivo methods and to compare it to the standard Transonics HD01 system in a clinical setting. MATERIALS AND METHODS: The Angioflow meter system consists of a 6-F endovascular catheter and a laptop computer containing proprietary software for this application. Bench-top testing with use of a flow model was performed to determine the accuracy of the Angioflow meter system. Initial in vivo studies were performed with use of an animal model to assess the endovascular performance of the Angioflow meter system. Subsequently, a human clinical trial was performed to compare the Angioflow meter to the standard Transonics HD01 system. Twenty-five patients with dysfunctional (<600 mL/min) hemodialysis grafts were referred for fistulography and angioplasty. Intragraft blood flow measurements were obtained before and after angioplasty with use of both the Angioflow meter system and the Transonics HD01 system. A comparison of the two systems was performed. RESULTS: Bench-top testing and animal studies demonstrated an excellent (r =.98) correlation between the measurements of the Angioflow meter and volumetric flow measurements. In the clinical trial, there was reasonable correlation (r =.72) between the blood flow measurements obtained with use of the Angioflow meter and Transonics HD01 system. The reproducibility of consecutive measurements with the Angioflow meter was excellent (r =.98). The mean increase in intragraft blood flow after angioplasty was 320 mL/min. CONCLUSION: The Angioflow meter is an accurate and reliable endovascular device for measuring intragraft blood flow during interventional procedures. Use of this catheter-based system should prove beneficial for quantifying the success of endovascular interventions, the assessment of arterial inflow, and identification of inconspicuous lesions.
机译:目的:本研究的目的是通过体外和体内方法评估Angioflow流量计系统的准确性和可靠性,并将其与临床环境中的标准Transonics HD01系统进行比较。材料与方法:Angioflow流量计系统由6-F血管内导管和一台膝上型计算机组成,该膝上型计算机包含用于此应用程序的专有软件。使用流量模型进行了台式测试,以确定Angioflow流量计系统的准确性。使用动物模型进行了初步的体内研究,以评估Angioflow仪表系统的血管内性能。随后,进行了一项人体临床试验,将Angioflow流量计与标准Transonics HD01系统进行了比较。 25名功能失调(<600 mL / min)血液透析移植物的患者被转诊作了瘘管造影和血管成形术。在血管成形术之前和之后都使用Angioflow Meter系统和Transonics HD01系统获得了移植物内血流测量值。比较了两个系统。结果:台式测试和动物研究表明,血管流量仪的测量值与体积流量测量值之间具有极好的相关性(r = .98)。在临床试验中,使用Angioflow流量计和Transonics HD01系统获得的血流测量值之间存在合理的相关性(r = .72)。使用Angioflow流量计进行连续测量的重现性极好(r = .98)。血管成形术后移植物内血流的平均增加为320 mL / min。结论:Angioflow流量计是一种准确可靠的血管内装置,用于在介入手术过程中测量移植物内的血流。这种基于导管的系统的使用应被证明有助于量化血管内干预的成功率,评估动脉血流以及识别不明显的病变。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号