摘要

The automated laser-based hematology analyzer Sysmex XT-2000iVTM providing a complete blood cell count (CBC) and 5-part differential has been introduced in large veterinary laboratories. The aim of the current study was to determine precision, linearity, and accuracy of the Sysmex analyzer. Reference method for the accuracy study was the laser-based hematology analyzer ADVIAReg. 2120. For evaluation of accuracy, consecutive fresh blood samples from healthy and diseased cats (n=216), dogs (n=314), and horses (n=174) were included. A low intra-assay coefficient of variation (CV) of approximately 1% was seen for the CBC except platelet count (PLT). An intra-assay CV ranging between 2% and 5.5% was evident for the differential count except for feline and equine monocytes (7.7%) and horse eosinophils (15.7%). Linearity was excellent for white blood cell count (WBC), hematocrit value, red blood cell count (RBC), and PLT. For all evaluated species, agreement was excellent for WBC and RBC, with Spearman rank correlation coefficients (rs) ranging from >0.99 to 0.98. Hematocrit value correlated excellently in cats and dogs, whereas for horses, a good correlation was evident. A good correlation between both analyzers was seen in feline and equine PLT (rs=0.89 and 0.92, respectively), whereas correlation was excellent for dogs (rs=0.93). Biases were close to 0 except for mean corpuscular hemoglobin concentration (4.11 to -7.25 mmol/l) and canine PLT (57x109/l). Overall, the performance of the Sysmex analyzer was excellent and compared favorably with the ADVIA analyzer.
机译:基于激光的自动血液分析仪Sysmex XT-2000 i V TM 可提供完整的血细胞计数(CBC)和5部分差异,已在大型兽医实验室中引入。当前研究的目的是确定Sysmex分析仪的精度,线性和准确性。精度研究的参考方法是基于激光的血液学分析仪ADVIAReg。 2120.为了评估准确性,从健康和患病的猫( n = 216),狗( n = 314)和马( n = 174)。除血小板计数(PLT)外,CBC的测定内变异系数(CV)低,约为1%。除猫和马单核细胞(7.7%)和马嗜酸性粒细胞(15.7%)外,测定内CV差异在2%至5.5%之间。线性对于白细胞计数(WBC),血细胞比容值,红细胞计数(RBC)和PLT极好。对于所有评估物种,WBC和RBC的一致性都很好,Spearman等级相关系数(r s )的范围在> 0.99至0.98之间。血细胞比容值在猫和狗中具有极好的相关性,而对于马,则有很好的相关性。猫和马的PLT中两个分析仪之间的相关性很好(r s = 0.89和0.92),而狗的相关性极佳(r s = 0.93) 。除平均红细胞血红蛋白浓度(4.11至-7.25 mmol / l)和犬PLT(57x10 9 / l)外,偏差均接近于0。总体而言,Sysmex分析仪的性能非常出色,与ADVIA分析仪相比具有优势。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号