Evaluation of automated hematology analyzer DYMIND DH76 compared to SYSMEX XN 1000 system

机译:与Sysmex XN 1000系统相比,自动血液分析仪DYMIND DH76评估

获取原文
获取外文期刊封面目录资料

摘要

Background DYMIND DH76 (DYMIND BIOTECH, China) is a new automated hematology system designed to provide CBC count, including a 5-part WBC differential count, and its analytical performance should be assessed before adoption for clinical use. Methods The analyzer was evaluated according to the International Council for Standardization in Haematology guideline. The purposes of this study were to assess its analytical performance in comparison to SYSMEX XN 1000 hematology analyzer currently used in our laboratory, as well as to compare the automated and manual WBC differential. Results Within-run precision in all concentration ranges was very good with coefficients of variation (CVs) between 0.02% and 2.5% except for platelets over 500×109/L (CV 9.5%). Within-batch imprecision showed CVs lower the declared deviation ranges. Accuracy (defined as trueness) was excellent for all CBC and white cell differential parameters, compared with the state of the art%. Linearity was confirmed with excellent regression coefficients (0.999-1.000), even in the lowest values, and carryover was ≤ 1%. Comparison between DYMIND DH76 and SYSMEX XN 1000 was also very good with correlation coefficients (R2) for WBC (1.000), RBC (0.999), hemoglobin (0.999) and PLT over 50×10 ~(9)/L (0.994) and R2 was lower but still acceptable (0.910) for PLT&50×10 ~(9)/L. R2 for neutrophils, lymphocytes, eosinophils, basophils, and monocytes were 0.974, 0.982, 0.957, 0.625, and 0.836, respectively, in the comparison between the manual and DYMIND DH76 automated differential WBC counts. Conclusions With excellent analytical performance and acceptable comparative analysis, DYMIND DH76 hematology analyser covered the predefined international standards and requirements and is fully appropriate for clinical application.
机译:背景技术Dymind DH76(Dymind Biotech,中国)是一种新的自动血液系统,旨在提供CBC计数,包括5份WBC差分计数,并且在采用临床用途之前应评估其分析性能。方法分析仪根据国际血液学指南标准化委员会评估。该研究的目的是评估其实验室目前使用的Sysmex XN 1000血液学分析仪的分析性能,以及比较自动化和手动WBC差异。结果在所有浓度范围内的运行型精度非常良好,变异系数在0.02%和2.5%之间,除血小板超过500×109 / L(CV 9.5%)。在批处理中显示CVS降低声明的偏差范围。与现有技术相比,所有CBC和白细胞差分参数都非常出色(定义为Truenes)。用优异的回归系数(0.999-1.000)确认线性度,即使在最低值中,仍然≤1%。 DYMIND DH76和Sysmex XN 1000之间的比较也非常良好,WBC(1.000),RBC(0.999),血红蛋白(0.999)和PLT超过50×10〜(9)/ L(0.994)和R2的相关系数(R2) PLT&amp的较低但仍然可接受(0.910); 50×10〜(9)/ l。对于中性粒细胞,淋巴细胞,嗜酸性粒细胞,嗜碱性粒细胞和单核细胞的R2分别为0.974,0.982,0.957,0.625和0.836,在手动和Dymind DH76自动化WBC计数之间的比较中。结论具有优异的分析性能和可接受的比较分析,Dymind DH76血液分析仪涵盖了预定义的国际标准和要求,并完全适用于临床应用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号