Development of a dielectric barrier discharge ion source for ambient mass spectrometry

机译:开发用于环境质谱的介电势垒放电离子源

获取原文
获取原文并翻译 | 示例
           

摘要

A new ion source based on dielectric barrier discharge was developed as an alternative ionization source for ambient mass spectrometry. The dielectric barrier discharge ionization source, termed as DBDI herein, was composed of a copper sheet electrode, a discharge electrode, and a piece of glass slide in between as dielectric barrier as well as sample plate. Stable low-temperature plasma was formed between the tip of the discharge electrode and the surface of glass slide when an alternating voltage was applied between the electrodes. Analytes deposited on the surface of the glass slide were clesorbed and ionized by the plasma and the ions were introduced to the mass spectrometer for mass analysis. The capability of this new ambient ion source was demonstrated with the analysis of 20 amino acids, which were deposited on the glass slide separately. Protonated molecular ions of [M + H](+) were observed for all. the amino acids except for L-arginine. This ion source was also used for a rapid discrimination of L-valine, L-proline, L-serine and L-alanine from their mixture. The limit of detection was 3.5 pmol for L-alanine using single-ion-monitoring (SIM). Relative standard deviation (RSD) was 5.78% for 17.5 nmol of L-alanine (n = 5). With the advantages of small size, simple configuration and ease operation at ambient conditions, the dielectric barrier discharge ion source would potentially be coupled to portable mass spectrometers.
机译:开发了一种基于介电势垒放电的新型离子源,作为环境质谱的替代电离源。介电屏障放电电离源在本文中称为DBDI,由铜片电极,放电电极和介于两者之间的一块玻璃载玻片作为介电屏障以及样品板组成。当在电极之间施加交流电压时,在放电电极的尖端和载玻片的表面之间形成稳定的低温等离子体。沉积在载玻片表面上的分析物被等离子体裂解并离子化,并将离子引入质谱仪进行质量分析。通过分析20种氨基酸(分别沉积在载玻片上)证明了这种新型环境离子源的功能。全部观察到[M + H](+)的质子化分子离子。 L-精氨酸以外的氨基酸。该离子源还用于从混合物中快速鉴别L-缬氨酸,L-脯氨酸,L-丝氨酸和L-丙氨酸。使用单离子监测(SIM)的L-丙氨酸检测极限为3.5 pmol。 17.5 nmol的L-丙氨酸的相对标准偏差(RSD)为5.78%(n = 5)。由于具有体积小,配置简单和在环境条件下易于操作的优点,电介质势垒放电离子源可能会耦合到便携式质谱仪。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号