首页> 外文期刊>Journal of perinatology: Official journal of the California Perinatal Association >Impact of warmed inhaled gas from the mechanical ventilator on esophageal temperature during therapeutic whole body hypothermia
【24h】

Impact of warmed inhaled gas from the mechanical ventilator on esophageal temperature during therapeutic whole body hypothermia

机译:机械通气治疗期间,来自机械呼吸机的热吸入气体对食管温度的影响

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVE: During therapeutic whole body cooling (WBC), the core temperature is usually monitored with an esophageal probe. Most infants receive mechanical ventilation while being cooled. As the temperature in the esophagus responds rapidly to changes in the ambient temperature, inhalation of warmed gas from ventilator may lead to overestimation of ventilated patients' actual temperature, causing automated cooling devices to overcool patients well below the desired set temperature targets. To determine if the esophageal temperature recordings during therapeutic WBC differ between ventilated and non-ventilated infants.
机译:目的:在治疗性全身冷却(WBC)期间,通常使用食管探针监测中心温度。大多数婴儿在降温时都会接受机械通气。由于食道中的温度对环境温度的变化做出快速响应,因此从呼吸机吸入温暖的气体可能导致对通风患者的实际温度的过高估计,从而导致自动冷却设备使患者过冷,使其远低于所需的设定温度目标。确定在通气婴儿和非通气婴儿之间进行治疗性WBC时食道温度记录是否不同。