首页> 外文期刊>Clinical oncology >Verification films: a study of the daily and weekly reproducibility of breast patient set-up in the START trial.
【24h】

Verification films: a study of the daily and weekly reproducibility of breast patient set-up in the START trial.

机译:验证影片:在START试验中对乳腺患者设置的每日和每周可重复性进行的研究。

获取原文
获取原文并翻译 | 示例
       

摘要

AIMS: Verification of patient position in radiotherapy by the use of portal images is a current practice in most radiotherapy departments. All patients within the Standardisation of Breast Radiotherapy Trial (START) were required to have on-treatment verification images. Measurements of maximum lung and breast depth from these were used to provide details for the case-report forms. This study was undertaken to validate these measurements, assess the reproducibility of patients' set-up within the START trial and to compare reproducibility on a day-to-day basis with reproducibility over a longer period. Analysis also included a subjective assessment of image quality. MATERIALS AND METHODS: Fifty-three patients from 29 departments were studied. Some centres used electronic portal imaging and others used film. Where film verification was the method of choice, information was collected about the type of film and cassette used. RESULTS: The correlation coefficients for inter- and intra-observer variability of breast and lung depth were 0.99. Comparing maximum breast depth on weekly and daily images, the median per cent standard deviation of breast depth was 4% and 3%, respectively. The mean standard deviation of breast depth for all patients, which gives an indication of the individual patient variability, was 5.5 +/- 2.2 mm for weekly and 3.9 +/- 1.5 mm for daily measurements. For lung depth, mean standard deviation was 6.8 +/- 2.5 mm for weekly and 6.8 +/- 2.8 mm for daily measurements. Images taken using Kodak X-omat V film were inferior to those taken with films in corresponding cassettes. CONCLUSION: Differences between lung depth measured in simulator and on the treatment unit were evident for many patients. There was little difference in the standard deviations of breast and lung depth for daily and weekly verification films on the treatment machine.
机译:目的:大多数门诊放射科目前正在通过使用门户图像来验证患者在放射治疗中的位置。乳腺放射治疗试验(START)标准化范围内的所有患者都必须具有治疗中的验证图像。从这些测量中获得的最大肺和乳房深度被用来提供病例报告表的详细信息。进行这项研究是为了验证这些测量结果,评估START试验中患者设置的可重复性,并将日常的可重复性与更长时期的可重复性进行比较。分析还包括对图像质量的主观评估。材料与方法:研究了29个科室的53例患者。一些中心使用电子门户成像,其他中心使用胶卷。在选择胶片验证作为选择方法的地方,收集了有关所用胶片和暗盒类型的信息。结果:观察者之间和观察者内部的乳房和肺部深度的相关系数为0.99。比较每周和每日图像上的最大乳房深度,乳房深度的标准偏差中位数分别为4%和3%。所有患者的乳房深度平均标准差(表明个体患者的变异性)为每周5.5 +/- 2.2 mm,每天测量3.9 +/- 1.5 mm。对于肺深,每周的平均标准偏差为6.8 +/- 2.5毫米,而每日测量的平均标准偏差为6.8 +/- 2.8毫米。使用柯达X-omat V胶片拍摄的图像不如使用相应暗盒中的胶片拍摄的图像。结论:对于许多患者,在模拟器和治疗单元上测得的肺深之间存在明显差异。在治疗机上每天和每周验证胶片的胸部和肺部深度的标准偏差几乎没有差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号