Numerous discarded ProFile GT, ProFile, and ProTaper nickel-titanium rotary instruments obtained from two graduate endodontic clinics were examined by scanning electron microscopy. These instruments had an unknown history of clinical use and had fractured or experienced considerable permanent torsional deformation without complete separation. The failure processes generally exhibited substantial ductile character, evidenced by a dimpled rupture fracture surface. Crack propagation at grain boundaries and cleavage surfaces indicative of transgranular fracture were observed for some specimens. It appeared that oxide particles from the manufacturing process served as nucleating sites for the microvoids, leading to dimpled rupture. A previously unreported fracture mode also was observed, in which crack propagation, approximately parallel to the local flute orientation, connected pitted regions on the surface. Combining present and previous scanning electron microscopy observations of clinically failed instruments, suggestions are offered for improving their fracture resistance.
机译:通过扫描电子显微镜检查了从两个研究生牙髓诊所获得的大量废弃的ProFile GT,ProFile和ProTaper镍钛旋转仪。这些器械的临床使用历史不明,没有完全分离就断裂或经历了相当大的永久性扭转变形。破坏过程通常表现出相当大的延展性,这可以通过破裂的破裂表面来证明。对于某些样品,观察到裂纹在晶界和解理表面的传播表明晶界断裂。似乎来自制造过程的氧化物颗粒充当微孔的成核位点,导致凹坑破裂。还观察到以前未报告的断裂模式,其中裂纹扩展(近似平行于局部长笛方向)连接了表面上的凹坑区域。结合当前和以前对临床失败器械的扫描电子显微镜观察,提出了改善其抗断裂性的建议。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号