This study examined used, discarded rotary nickel-titanium instruments obtained from 14 endodontists in four countries, and identified factors that may influence defects produced during clinical use. A total of 7,159 instruments were examined for the presence of defects. Unwinding occurred in 12% of instruments and fractures in 5% (1.5% torsional, 3.5% flexural). The defect rates varied significantly among endodontists. Instrument design factors also influenced defect rate, but to a lesser extent. The mean number of uses of instruments with and without defects was 3.3 +/- 1.8 (range: 1-10), and 4.5 +/- 2.0 (range: 1-16), respectively. The most important influence on defect rates was the operator, which may be related to clinical skill or a conscious decision to use instruments a specified number of times or until defects were evident.
机译:这项研究检查了从四个国家的14位牙髓医生那里获得的二手,废弃的旋转镍钛仪器,并确定了可能影响临床使用过程中产生的缺陷的因素。总共检查了7,159台仪器是否存在缺陷。退绕发生在12%的器械中,骨折发生在5%(扭转1.5%,弯曲3.5%)。牙髓病患者的缺陷率差异很大。仪器设计因素也影响缺陷率,但程度较小。有缺陷器械和无缺陷器械的平均使用次数分别为3.3 +/- 1.8(范围:1-10)和4.5 +/- 2.0(范围:1-16)。对缺陷率的最重要影响是操作员,这可能与临床技能或有意识地决定使用仪器指定次数或直到发现明显缺陷有关。