State-of-the-art vitamin D assays: A comparison of automated immunoassays with liquid chromatography-tandem mass spectrometry methods

机译:最先进的维生素D测定:液相色谱-串联质谱法与自动免疫测定的比较

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: Vitamin D testing is increasing worldwide. Recently several diagnostic manufacturers including Abbott and Siemens have launched automated 25-hydroxy vitaminD(25OH-D) immunoassays. Furthermore, preexisting assays from DiaSorin and Roche have recently been modified. We compared the performance of 5 automated immunoassays, an RIA and 2 liquid chromatography-tandem mass spectrometry (LC-MS/MS) methods. METHODS: Aliquots of 170 randomly selected patient samples were prepared and 25OH-D was measured by 2 LC-MS/MS methods, an RIA (DiaSorin), and automated immunoassays from Abbott (Architect), DiaSorin (LIAISON), IDS (ISYS), Roche (E170, monoclonal 25OH-D 3 assay), and Siemens (Centaur). Within-run and between-run imprecision were evaluated by measurement of 5 replicates of 2 serum pools on 5 consecutive days. RESULTS: The LC-MS/MS methods agreed, with a concordance correlation coefficient (CCC) of 0.99 and bias of 0.56 μg/L (1.4 nmol/L). The RIA assay showed a performance comparable to LC-MS/MS, with aCCCof 0.97 and a mean bias of 1.1 μg/L (2.7 nmo/L). All immunoassays measured total 25OH-D (including D 3and D 2), with the exception of the Roche assay (D 3only). Among the immunoassays detecting total 25OH-D, the CCCs varied between 0.85 (Abbott) to 0.95 (LIAISON). The mean bias ranged between 0.2 μg/L (0.5 nmol/L) (LIAISON) and 4.56 μg/L (11.4 nmol/L) (Abbott). The Roche 25OH-D 3 assay demonstrated small mean bias [-2.7 μg/L (-6.7 nmol/L)] [-2.7 μg/L (-6.7 nmol/L)] but a low CCC of just 0.66. Most assays demonstrated good intra- and interassay precision, with CV 10%. CONCLUSIONS: Automated immunoassays demonstrated variable performance and not all tests met our minimum performance goals. It is important that laboratories be aware of the limitations of their assay.
机译:背景:维生素D测试在世界范围内正在增加。最近,包括雅培(Abbott)和西门子(Siemens)在内的数家诊断制造商已启动了自动25-羟基维生素D(25OH-D)免疫测定。此外,最近修改了DiaSorin和Roche的现有检测方法。我们比较了5种自动免疫分析,RIA和2种液相色谱-串联质谱分析(LC-MS / MS)方法的性能。方法:准备了170个随机选择的患者样品的等分试样,并通过2种LC-MS / MS方法,RIA(DiaSorin)和来自Abbott(Architect),DiaSorin(LIAISON),IDS(ISYS)的自动免疫测定法测量了25OH-D ,罗氏(E170,单克隆25OH-D 3分析)和西门子(半人马座)。通过连续5天测量2个血清库的5个重复样本来评估组内和组间不准确性。结果:LC-MS / MS方法一致,一致性相关系数(CCC)为0.99,偏差为0.56μg/ L(1.4 nmol / L)。 RIA分析显示出与LC-MS / MS相当的性能,aCCC为0.97,平均偏差为1.1μg/ L(2.7 nmo / L)。除罗氏测定法(仅D 3)外,所有免疫测定法均测量了总的25OH-D(包括D 3和D 2)。在检测总25OH-D的免疫测定中,CCC在0.85(Abbott)至0.95(LIAISON)之间变化。平均偏差介于0.2μg/ L(0.5 nmol / L)(LIAISON)和4.56μg/ L(11.4 nmol / L)(Abbott)之间。 Roche 25OH-D 3分析显示出较小的平均偏差[-2.7μg/ L(-6.7 nmol / L)] [-2.7μg/ L(-6.7 nmol / L)],但CCC仅为0.66。大多数测定法显示出良好的测定内和测定间精密度,CV <10%。结论:自动化免疫分析显示出可变的性能,并且并非所有测试都达到了我们的最低性能目标。重要的是,实验室必须意识到其检测方法的局限性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号