USE OF A THERMALLY INSULATED COLUMN FOR IMPROVED SPEED, EFFICIENCY AND RESOLUTION IN PACKED-COLUMN SUPERCRITICAL FLUID CHROMATOGRAPHY

机译:热绝缘柱在填充柱超临界流体色谱中提高速度,效率和分辨率的用途

获取原文
获取原文并翻译 | 示例

摘要

The use of a thermally insulated column for significant improvements in speed, efficiency and resolution in packed-column supercritical fluid chromatography at intermediate mobile: phase densities is demonstrated. Elution of a mixture of normal alkanes, containing 12 to 18 carbon atoms from a 150x2.0-mm I.D. column packed with 5-mu m porous octylsilica particles, was performed with carbon dioxide mobile phase at 50 degrees C at reduced densities of 1.0 and 1.5. For elution of n-octadecane at a reduced density of 1.0, a two-to sixteen-fold increase in the apparent number of theoretical plates was observed for an insulated column compared to the uninsulated column at pressure drops of 12 and 29 bar, respectively, along with decreased retention times and improved resolution. The improved efficiency obtained with an insulated column at the lower density is due to the effective elimination of radial temperature gradients associated with expansion and cooling of the mobile phase. The insulated column provided no measurable improvement at the higher density. (C) 1997 Elsevier Science B.V
机译:展示了使用隔热柱显着改善填充柱超临界流体色谱在中间流动相下的速度,效率和分离度:从150x2.0-mm内径洗脱含12至18个碳原子的正构烷烃混合物。在50℃下以1.0和1.5的降低的密度,用二氧化碳流动相在装有5微米多孔辛基二氧化硅颗粒的色谱柱上进行色谱分离。为了以1.0的降低密度洗脱正十八烷,在压力降分别为12和29 bar的情况下,与未绝缘的色谱柱相比,与未绝缘的色谱柱相比,观察到的理论塔板表观数量增加了2到16倍,以及减少的保留时间和提高的分离度。用较低密度的绝缘柱获得的效率提高是由于有效消除了与流动相膨胀和冷却相关的径向温度梯度。绝缘柱在较高密度下没有提供可测量的改进。 (C)1997年Elsevier Science B.V

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号