首页> 外文期刊>Journal of cataract and refractive surgery >Protecting the haptic when a Monarch injector is used.
【24h】

Protecting the haptic when a Monarch injector is used.

机译:使用Monarch注射器时保护触觉。

获取原文
获取原文并翻译 | 示例

摘要

Tsuneoka and coauthors'1 summary of their experience using the AcrySof intraocular lens (IOL) with a Monarch injector is well written and of great relevance. We, too, have found this to be a useful technique and wanted to share a point that may not have been stressed enough in this article.When placing the 6.5 mm MA50 IOL into the Monarch injector, it is important to ensure that the trailing haptic is out of the way of the plunger. The Monarch II B cartridge has a peg, which allows the surgeon to set the trailing haptic off to the side (Figure 1). This allows the plunger to go forward, pushing the optic and avoiding the haptic. As the IOL goes into the cartridge, the haptic will flip and go over the plunger, allowing safe delivery into the eye. If the haptic is not put to the side and protected in this manner, the haptic will be damaged by the plunger, something that occurred at our institution in the past.
机译:Tsuneoka和合著者1撰写的有关使用Monarch注射器使用AcrySof人工晶状体(IOL)的经验的总结很恰当,并且具有很大的相关性。我们也发现这是一种有用的技术,并且希望分享本文可能没有足够强调的观点。将6.5 mm MA50 IOL放入Monarch注射器中时,重要的是要确保尾随触觉挡住了柱塞。 Monarch II B弹药筒有一个钉子,使外科医生可以将尾随触觉偏向侧面(图1)。这样可以使柱塞向前移动,推动光学元件并避开触觉。当IOL进入药筒时,the将翻转并越过柱塞,从而可以安全地输送到眼睛中。如果未将触觉器放在一边并以这种方式进行保护,则触觉器将受到柱塞的损坏,这是我们机构过去发生的事情。