Optical and ultrasound measurement of axial length and anterior chamber depth for intraocular lens power calculation.

机译:眼轴长度和前房深度的光学和超声测量,用于计算人工晶状体度数。

获取原文
获取原文并翻译 | 示例
           

摘要

To evaluate the precision, reproducibility, and applicability of an optical method based on partial coherence interferometry for intraocular lens (IOL) power calculation.Ultrasound laboratory of a university eye hospital.A prospective comparison of measurements made by the IOLMaster optical instrument (Carl Zeiss) and Ultrascan Digital 2000 contact ultrasound A-scan (Alcon) for IOL calculations was performed. Examined were 255 eyes of 134 persons (204 phakic, 47 pseudophakic, and 4 aphakic). The mean age of the patients was 67.9 years (range 7 to 94 years).The IOLMaster measurements were successful in more than 80% of cases: in 82%, 99%, and 99% for axial length (AL), anterior chamber depth (ACD), and keratometry measurements, respectively. The reproducibility of the AL and ACD measurements was very high (coefficient of variation 0.13% and 2.20%, respectively). The AL and ACD values were significantly larger with the IOLMaster (P <.001) than with the Ultrascan Digital 2000. The correlation between ultrasound and optical AL measurements was high (r = 0.985; P <.001); however, there was no correlation between ACD measurements (r = 0.079; P =.397). The corneal refractive power measurements of a Javal-type keratometer and the IOLMaster were highly correlated (r = 0.955; P <.001), with a mean difference of 0.2 diopter.The results show that measurements for IOL calculation are easy and precise with the optical method. It is a noncontact method, so no anesthesia is needed and there is no risk of infection.
机译:评估基于部分相干干涉法的人工晶状体(IOL)屈光度计算光学方法的精度,可重复性和适用性大学眼科医院的超声实验室对IOLMaster光学仪器(Carl Zeiss)进行的测量的前瞻性比较然后进行用于IOL计算的Ultrascan Digital 2000接触超声A扫描(Alcon)。检查了134人的255眼(204晶状体,47假晶状体和4无晶状体)。患者的平均年龄为67.9岁(范围为7至94岁)。IOLMaster测量在80%以上的病例中获得成功:轴向长度(AL),前房深度分别为82%,99%和99% (ACD)和角膜曲率测量。 AL和ACD测量的可重复性非常高(变异系数分别为0.13%和2.20%)。使用IOLMaster的AL和ACD值明显大于使用Ultrascan Digital 2000的AL和ACD值(P <.001)。超声和光学AL测量值之间的相关性很高(r = 0.985; P <.001)。但是,ACD测量值之间没有相关性(r = 0.079; P = 0.397)。 Javal型角膜曲率计和IOLMaster的角膜屈光力测量值高度相关(r = 0.955; P <.001),平均差为0.2屈光度。光学方法。这是一种非接触式方法,因此无需麻醉,也没有感染的风险。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号