Mechanical comparison of biodegradable femoral fixation devices for hamstring tendon graft--a biomechanical study in a porcine model.

机译:绳肌腱移植的可生物降解股骨固定装置的机械比较-猪模型中的生物力学研究。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: Initial fixation strength is critical for the early post-operative rehabilitation of patients with anterior cruciate ligament reconstructions. However, even the best femoral fixation devices remain controversial. We compared the biomechanical characteristics of tendon grafts fixed by different biodegradable femoral fixation devices following anterior cruciate ligament reconstruction. METHODS: The Bio-TransFix, Rigidfix, Bioscrew with EndoPearl augmentation and Bioscrew devices were used to fix porcine flexor digitorum profundus tendon grafts in 32 porcine femora. Displacement of each tendon graft was evaluated after cyclic loading testing. Stiffness, ultimate failure load and failure mode of these fixation devices were measured with load-to-failure testing. FINDINGS: The displacement of the femur-graft-cement complex in response to cyclic loading was lower (P<0.05) for the Bio-TransFix than the Rigidfix, Bioscrew with EndoPearl augmentation, and Bioscrew groups. The fixation stiffness values of the Rigidfix and the Bioscrew were significantly greater (P<0.05) than that of the Bio-TransFix. The ultimate failure load was significantly greater for the Bio-TransFix and the Rigidfix than the Bioscrew with EndoPearl augmentation or the Bioscrew (P<0.05). INTERPRETATION: The Bio-TransFix provided the least graft displacement under cyclic loading. However, this device gave less stability. The Rigidfix device provided better stability and stiffness of the tendon graft among those fixation devices that showed no significant differences in graft displacement under cyclic loading. However, no single fixation device provided less displacement along with a larger failure load and stiffness in this study.
机译:背景:初始固定强度对于前交叉韧带重建患者的术后早期康复至关重要。但是,即使是最好的股骨固定装置仍然存在争议。我们比较了前交叉韧带重建后不同生物可降解股骨固定装置固定的腱移植物的生物力学特性。方法:使用Bio-TransFix,Rigidfix,带EndoPearl增强的Bioscrew和Bioscrew装置将猪屈指趾深部肌腱移植物固定在32头猪股骨中。循环载荷测试后评估每个腱移植物的位移。这些固定装置的刚度,最终失效载荷和失效模式通过载荷-失效测试进行了测量。结果:Bio-TransFix的股骨-骨水泥结合体复合物响应循环负荷的位移低于Rigidfix,EndoPearl增强的Bioscrew和Bioscrew组(P <0.05)。 Rigidfix和Bioscrew的固定刚度值明显高于Bio-TransFix(P <0.05)。与带有EndoPearl增强的Bioscrew或Bioscrew相比,Bio-TransFix和Rigidfix的最终破坏载荷明显更大(P <0.05)。解释:在循环载荷下,Bio-TransFix提供了最少的移植物移位。但是,该装置稳定性较差。在那些在循环载荷下移植物移位没有明显差异的固定装置中,Rigidfix装置提供了更好的稳定性和刚度。但是,在这项研究中,没有单个固定装置可提供较小的位移以及较大的破坏载荷和刚度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号