首页> 外文期刊>Journal of clinical monitoring and computing >Safe transcranial electric stimulation motor evoked potential monitoring during posterior spinal fusion in two patients with cochlear implants
【24h】

Safe transcranial electric stimulation motor evoked potential monitoring during posterior spinal fusion in two patients with cochlear implants

机译:两名人工耳蜗植入患者后路脊柱融合术中安全经颅电刺激运动诱发电位监测

获取原文
获取原文并翻译 | 示例
           

摘要

Transcranial electric stimulation (TES) motor evoked potentials (MEPs) have become a regular part of intraoperative neurophysiologic monitoring (IONM) for posterior spinal fusion (PSF) surgery. Almost all of the relative contraindications to TES have come and gone. One exception is in the case of patients with a cochlear implant (CI). Herein we illustrate two cases of pediatric patients with CIs who underwent PSF using TES MEPs as part of IONM. In both instances the patients displayed no untoward effects from TES, and post-operatively both CIs were intact and functioning as they were prior to surgery.
机译:经颅电刺激(TES)运动诱发电位(MEP)已成为后路脊柱融合(PSF)手术的术中神经生理监测(IONM)的常规部分。 TES的几乎所有禁忌症都来了又去了。人工耳蜗(CI)的患者是一个例外。在此,我们说明了2例使用TES MEP作为IONM一部分进行PSF的小儿CI患者。在这两种情况下,患者均未表现出TES的不良影响,并且术后两个CI均完好无损,并且功能均与手术前相同。