首页> 外文期刊>Journal of biomechanical engineering. >Assessment of the Ocular Response Analyzer as a Tool for Intraocular Pressure Measurement
【24h】

Assessment of the Ocular Response Analyzer as a Tool for Intraocular Pressure Measurement

机译:眼反应分析仪作为眼压测量工具的评估

获取原文
获取原文并翻译 | 示例

摘要

The ocular response analyzer (ORA) is a new indentation tonometer that subjects the cornea to an increasing then decreasing air pulse, and uses the corresponding two applanation pressures P_1 and P_2 to estimate the intraocular pressure (IOP). The present study aims to improve the accuracy of IOP estimation through representative numerical simulation of the ORA procedure. A parametric study has been carried out to consider the effect of corneal thickness, curvature, age, and true IOP on the P_1 and P_2 measurements. Based on the obtained database of input and output parameters, an equation has been suggested relating the P_1 and P_2 measurements to the value of IOP. The equation is intended to make IOP estimates independent of corneal stiffness, which varies with size and age. The equation has been tested against a clinical data set obtained at Moorfields Eye Hospital involving 144 subjects, and found to produce IOP estimates that are less dependent on corneal size and age than published equations.
机译:眼反应分析仪(ORA)是一种新的压痕眼压计,它使角膜经受增大然后减小的空气脉冲,并使用相应的两个压平压力P_1和P_2来估算眼内压(IOP)。本研究旨在通过ORA程序的代表性数值模拟来提高IOP估计的准确性。已经进行了参数研究,以考虑角膜厚度,曲率,年龄和真实IOP对P_1和P_2测量的影响。根据获得的输入和输出参数数据库,提出了一个方程,将P_1和P_2测量值与IOP值相关。该方程式旨在使IOP估计不受角膜刚度的影响,后者随大小和年龄的变化而变化。已针对Moorfields眼科医院的144名受试者的临床数据集对该方程进行了测试,发现该方程产生的IOP估计值比已发表的方程式对角膜大小和年龄的依赖性更小。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号