The objective of this work was to develop a robotic device to perform biopsy and therapeutic interventions in the breast with real-time magnetic resonance imaging (MRI) guidance. The device was designed to allow for (i) stabilization of the breast by compression, (ii) definition of the interventional probe trajectory by setting the height and pitch of a probe insertion apparatus, and (iii) positioning of an interventional probe by setting the depth of insertion. The apparatus is fitted with five computer-controlled degrees of freedom for delivering an interventional procedure. The entire device is constructed of MR compatible materials, i.e. nonmagnetic and non-conductive, to eliminate artifacts and distortion of the MR images. The apparatus is remotely controlled by means of ultrasonic motors and a graphical user interface, providing real-time MR-guided planning and monitoring of the operation. Joint motion measurements found probe placement in less than 50 s and sub-millimeter repeatability of the probe tip for same-direction point-to-point movements. However, backlash in the rotation joint may incur probe tip positional errors of up to 5 mm at a distance of 40 mm from the rotation axis, which may occur for women with large breasts. The imprecision caused by this backlash becomes negligible as the probe tip nears the rotation axis. Real-time MR-guidance will allow the physician to correct this error Compatibility of the device within the MR environment was successfully tested on a 4 Tesla MR human scanner
机译:这项工作的目的是开发一种机器人设备,以实时磁共振成像(MRI)指导对乳房进行活检和治疗。该设备的设计允许(i)通过压缩来稳定乳房,(ii)通过设置探针插入设备的高度和间距来定义介入探针的轨迹,以及(iii)通过设定探针的位置来确定介入探针的位置。插入深度。该设备配有五个计算机控制的自由度,用于执行介入程序。整个设备由MR兼容材料制成,即非磁性和非导电性,以消除MR图像的伪影和失真。该设备通过超声波马达和图形用户界面进行远程控制,提供实时的MR引导计划和操作监控。联合运动测量发现,在不到50秒的时间内即可放置探头,并且对于相同方向的点对点移动,探头尖端的亚毫米可重复性。但是,旋转接头中的间隙可能会在距旋转轴40毫米的距离处引起高达5毫米的探针尖端位置误差,这可能对于乳房较大的女性会发生。当探针靠近旋转轴时,由该间隙引起的不精确度可以忽略不计。实时MR指导可让医生纠正此错误。MR环境中设备的兼容性已在4 Tesla MR人类扫描仪上成功测试

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号