摘要

A new modification of electrical low pressure impactor (ELPI) for the particle effective density measurement is presented. The system is capable of real-time operation and it is based on the serial measurement of mobility and aerodynamic diameter. In the studied configuration, a zeroth order mobility analyser is installed inside of the ELPI-instrument. The system is feasible for single modal distributions. For several particle materials and varying size distributions, the measured average density values were within 15% of the values obtained with a reference method.
机译:提出了一种新型的低压电气冲击器(ELPI),用于颗粒有效密度测量。该系统能够实时运行,并且基于对机动性和空气动力学直径的连续测量。在研究的配置中,ELPI仪器内部安装了零阶迁移率分析仪。该系统对于单模态分布是可行的。对于几种颗粒材料和变化的尺寸分布,测得的平均密度值在参考方法获得的值的15%以内。