Speciation of Manganese Using a Pneumatic Flow Injection Analysis-Tandem Spectrometer System

机译:气动流动注射-串联质谱仪系统分析锰的形态

获取原文
获取原文并翻译 | 示例

摘要

A pneumatic flow injection-tandem spectrometer system,without a delivery pump,was used for the speciation of manganese(Mn).In this system the suction force of the pneumatic nebulizer of a flame atomic absorption spectrometer(FAAS)was used for solution delivery through the manifold.Mn(VII)and total Mn(Mn(VII)and Mn(II))concentrations were determined using a UV-Vis spectrometer and FAAS,respectively.The Mn(II)concentration was determined by the difference between the two.The calibration curves were linear up to 15.00 mg L~(-1)for Mn(VII)and total Mn,with a detection limit of 0.08 mg L~(-1)and 0.05 mg L~(-1)for Mn(VII)and Mn,respectively.The mid-range precision and accuracy were < 1.89% and ± 2.50% for the 2 species,respectively,at a sampling rate of 80 samplings h~(-1).This system was used for the determination of Mn(VII)and Mn(II)in spiked and natural water,as well as industrial water.
机译:使用不带输送泵的气动注射串联质谱仪系统来形成锰(Mn)的形态。在该系统中,火焰原子吸收光谱仪(FAAS)的气动雾化器的吸力用于通过分别使用UV-Vis光谱仪和FAAS测定Mn(VII)以及总Mn(Mn(VII)和Mn(II))浓度。通过两者之间的差异确定Mn(II)浓度。 Mn(VII)和总Mn的校准曲线线性至15.00 mg L〜(-1),Mn(VII)的检出限为0.08 mg L〜(-1)和0.05 mg L〜(-1) )和Mn。在80个采样h〜(-1)的采样率下,这2种的中位精度和准确度分别为<1.89%和±2.50%。加标和天然水中以及工业用水中的Mn(VII)和Mn(II)。