Platelet capacity of various platelet pooling systems for buffy coat-derived platelet concentrates.

机译:血沉棕黄层衍生的血小板浓缩液的各种血小板合并系统的血小板容量。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

BACKGROUND: Platelet (PLT) storage lesions might depend on the total PLT count in the storage container and also on the PLT pooling system, especially the storage container, that is used for preparation of PLT concentrates (PCs). In this study, the PLT capacity of four commercially available PLT pooling systems was studied. MATERIALS AND METHODS: Four PCs were prepared in pooling systems of Baxter, Fresenius, Terumo, or Pall. The PCs were pooled and divided with various total PLT counts over the four storage containers (<225 x 10(9), 225 x 10(9)-324 x 10(9), 325 x 10(9)-424 x 10(9), and >424 x 10(9) PLTs). Volumes were kept equal by adding plasma to PCs with less than 425 x 10(9) PLTs until a same volume as for PCs with more than 424 x 10(9) PLTs was reached. PCs were stored at room temperature and tested for various in vitro variables on Days 1, 3, 5, 7, and 9. Paired experiments were repeated for each system five times. RESULTS: In vitro variables remained good for 9 days, that is, swirling score of 2 or more, pH value of 6.8 or more, glucose level of 10 mmol per L or more, lactate level of less than 25 mmol per L, and CD62p expression of less than 50 percent, for PCs in Baxter systems with more than 225 x 10(9) PLTs, for PCs in Fresenius and Terumo systems with 225 x 10(9) to 424 x 10(9) PLTs, and for PCs in Pall systems with fewer than 425 x 10(9) PLTs. CONCLUSION: PLT capacity depended on the PLT pooling systems used. All systems provide acceptable storage conditions. The Baxter system was the only system with capacity for more than 424 x 10(9) PLTs per PC.
机译:背景:血小板(PLT)储存损伤可能取决于储存容器中的总PLT计数,还取决于用于制备PLT浓缩物(PC)的PLT汇集系统,尤其是储存容器。在这项研究中,研究了四种市售PLT合并系统的PLT容量。材料与方法:在Baxter,Fresenius,Terumo或Pall的合并系统中准备了四台PC。将PC汇集起来,并在四个存储容器中分别分配不同的PLT总数(<225 x 10(9),225 x 10(9)-324 x 10(9),325 x 10(9)-424 x 10( 9)和> 424 x 10(9)PLT)。通过将血浆添加到PLT小于425 x 10(9)的PC,直到达到与体积大于424 x 10(9)的PC相同的体积来保持体积相等。将PC储存在室温下,并在第1、3、5、7和9天测试各种体外变量。每个系统重复配对实验五次。结果:体外变量连续9天保持良好,即涡旋得分为2或更高,pH值为6.8或更高,葡萄糖水平为10 L / L或更高,乳酸水平低于25 mmol / L和CD62p对于具有225 x 10(9)PLT的Baxter系统中的PC,具有225 x 10(9)至424 x 10(9)PLT的Fresenius和Terumo系统中的PC以及具有少于425 x 10(9)个PLT的颇尔系统。结论:PLT容量取决于所使用的PLT池系统。所有系统都提供可接受的存储条件。百特系统是唯一每台PC容量超过424 x 10(9)PLT的系统。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号