This study sought to quantitatively evaluate the reduction of radiation dose in coronary angiography and angioplasty with the use of image noise reduction technology in a routine clinical setting. Radiation dose data from consecutive 605 coronary procedures (397 consecutive coronary angiograms and 208 consecutive coronary interventions) performed from October 2014 to April 2015 on a coronary angiography system with noise reduction technology (Allura Clarity IQ) were collected. For comparison, radiation dose data from consecutive 695 coronary procedures (435 coronary angiograms and 260 coronary interventions) performed on a conventional coronary angiography system from October 2013 to April 2014 were evaluated. Patient radiation dosage was evaluated based on the cumulative dose area product. Operators and operator practice did not change between the 2 evaluated periods. Patient characteristics were collected to evaluate similarity of patient groups. Image quality was evaluated on a 5-grade scale in 30 patients of each group. There were no significant differences between the 2 evaluated groups in gender, age, weight, and fluoroscopy time (6.8 +/- 6.1 vs 6.9 +/- 6.3 minutes, not significant). The dose area product was reduced from 3195 +/- 2359 to 983 +/- 972 cGycm(2) (65%, p <0.001) in coronary angiograms and from 7123 +/- 4551 to 2431 +/- 1788 cGycm(2) (69%, p <0.001) in coronary interventions using the new noise reduction technology. Image quality was graded as similar between the evaluated systems (4.0 +/- 0.7 vs 4.2 +/- 0.6, not significant). In conclusion, a new x-ray technology with image noise reduction algorithm provides a substantial reduction in radiation exposure without the need to prolong the procedure or fluoroscopy time. (C) 2016 Elsevier Inc. All rights reserved.
机译:这项研究试图在常规临床环境中使用图像降噪技术定量评估冠状动脉造影和血管成形术中放射剂量的减少。收集了2014年10月至2015年4月在具有降噪技术(Allura Clarity IQ)的冠状动脉造影系统上执行的连续605次冠状动脉手术(397次连续冠状动脉造影和208次连续冠状动脉介入治疗)的辐射剂量数据。为了进行比较,我们评估了从2013年10月至2014年4月在常规冠状动脉造影系统上执行的连续695例冠状动脉手术(435例冠状动脉造影和260例冠状动脉介入治疗)的辐射剂量数据。根据累积剂量面积积评估患者的放射剂量。操作人员和操作人员的做法在两个评估期间之间没有变化。收集患者特征以评估患者组的相似性。在每组的30位患者中以5级等级评估图像质量。 2个评估组在性别,年龄,体重和透视检查时间之间无显着差异(6.8 +/- 6.1分钟与6.9 +/- 6.3分钟,无显着差异)。剂量面积乘积在冠状动脉造影中从3195 +/- 2359降低到983 +/- 972 cGycm(2)(65%,p <0.001)从7123 +/- 4551降低到2431 +/- 1788 cGycm(2)新型降噪技术在冠状动脉介入治疗中的成功率为(69%,p <0.001)。评估系统之间的图像质量评级为相似(4.0 +/- 0.7与4.2 +/- 0.6,不显着)。总而言之,一种具有图像降噪算法的新型X射线技术可大幅减少辐射暴露,而无需延长程序或荧光检查时间。 (C)2016 Elsevier Inc.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号