摘要

We describe a dislocation after Scorpio mobile-bearing total knee arthroplasty. This system is a rotating platform posterior-stabilized design and utilizes a single post as part of the metal tibial tray. Only one locking ring inside the socket of the polyethylene insert secures a tight connection with the post. Spontaneous dislocation between the polyethylene insert and the metal tray occurred at 22 months post surgery while rising from the supine position with slight knee flexion. Operative findings revealed failure of the locking ring and the original insert was replaced with a thicker insert. Our case and a duplicated saw bone model demonstrated that failure of the locking system resulted in the dislocation of the insert.
机译:我们描述了天蝎座活动全膝关节置换术后的脱位。该系统是旋转平台后稳定设计,并利用单个支柱作为金属胫骨托的一部分。聚乙烯插件插槽内只有一个锁紧环可确保与立柱牢固连接。聚乙烯插入物和金属托之间的自发脱位发生在术后22个月,同时从仰卧位上升,膝盖略有屈曲。手术发现发现锁紧环发生故障,原来的刀片被较厚的刀片替换。我们的病例和重复的锯骨模型表明,锁定系统的故障导致刀片移位。