首页> 外文期刊>The Journal of arthroplasty >Secondary knee instability caused by fracture of the stabilizing insert in a dual-articular total knee.
【24h】

Secondary knee instability caused by fracture of the stabilizing insert in a dual-articular total knee.

机译:由双关节全膝关节中稳定插入物断裂引起的继发性膝关节不稳定性。

获取原文
获取原文并翻译 | 示例
           

摘要

A case of a fractured polyethylene stabilizing insert causing secondary knee instability in a Dual-articular total knee arthroplasty (TKA) is presented. A 65-year-old woman who underwent surgery with a Dual-articular TKA 4 years earlier had a well-functioning prosthesis until a fall, after which she began complaining of an intermittent, audible clicking; pain; and instability. At surgery, a fractured polyethylene peg was discovered, and the tibial insert was replaced. At follow-up, she remains asymptomatic. This is the first description of a fractured stabilizing insert in a Dual-articular knee.
机译:提出了一个骨折的聚乙烯稳定插入物引起双关节全膝关节置换术(TKA)中继发性膝关节不稳的病例。一名65岁的妇女在4年前接受了双关节TKA手术后,其假体功能良好,直到跌倒为止,之后她开始抱怨间歇性的可听见的喀哒声。痛;和不稳定。在手术中,发现骨折的聚乙烯钉,并替换了胫骨插入物。在随访中,她没有症状。这是双关节膝盖中稳定的骨折插入物的首次描述。