The purpose of this study was to assess the durability of a non-cemented, hemispherical rim-fit, hydroxyapatite coated cup with a highly cross-linked polyethylene in 223 total hip arthroplasties. At 6-years follow-up (range, 5-9), there were no cup revisions for osteolysis or loosening. Radiologic evidence of osseointegration was based on presence of Stress Induced Reactive Cancellous Bone and radial trabeculae, seen in 47% and 93% of cups, respectively; both were most prevalent in Zone 1. There was no interference demarcation in any zones. Two cups were revised (0.9%): one for dislocation and another for infection. The Kaplan-Meier survivorship for cup revision for any failure (infection, dislocation) was 99% and for mechanical failure (osteolysis, loosening) was 100%. This design has excellent safety, efficacy and durability. (C) 2015 Elsevier Inc. All rights reserved.
机译:这项研究的目的是评估具有高交联聚乙烯的非胶结,半球形边缘贴合,羟基磷灰石涂层的杯子在223例全髋关节置换术中的耐久性。在6年的随访中(范围5-9),没有因骨质溶解或松弛而变杯的情况。骨整合的放射学证据是基于压力诱发的反应性松质骨和radial骨小梁的存在,分别在47%和93%的杯子中可见;两者都在区域1中最普遍。在任何区域中都没有干扰划分。修改了两杯(0.9%):一个用于脱臼,另一个用于感染。对于任何失败(感染,脱位)进行杯翻修的Kaplan-Meier存活率为99%,而对于机械失败(溶骨,松动)为100%。这种设计具有出色的安全性,功效和耐用性。 (C)2015 Elsevier Inc.保留所有权利。