We conducted a study to determine if there is a significant difference in ultimate load, yield strength, and fatigue strength between solid core screws and cannulated screws and if a solid core insert placed inside a cannulated screw would have biomechanical effects similar to those of a solid core screw. Five screw designs were tested: Synthes 4.5-mm solid core and 4.5-mm cannulated and our prototype 4.0-mm solid core, 4.0-mm cannulated, and 4.0-mm cannulated with solid core insert. Biomechanical testing with 3-point bending was used to determine ultimate load, yield strength, and cycles to failure for 6 screws of each design. Ultimate load, yield strength, and cycles to failure were significantly (P < .05) lower in the Synthes 4.5-mm cannulated screws than in the Synthes solid core screws and significantly lower in the prototype cannulated screws than in the prototype solid core screws (P < .05) and prototype cannulated screws with solid core inserts (P < .05). There was no significant difference (P > .05) in ultimate load, yield strength, or cycles to failure between the prototype cannulated screws with solid core inserts and the prototype solid core screws.
机译:我们进行了一项研究,以确定实心螺钉和空心螺钉之间的极限载荷,屈服强度和疲劳强度是否存在显着差异,以及置入空心螺钉内部的实心插入件是否具有与实心螺钉相似的生物力学效果。核心螺丝。测试了五种螺钉设计:Synthes 4.5毫米实心铁芯和4.5毫米空心针,以及我们的原型4.0毫米实心铁芯,4.0毫米空心针和4.0毫米空心针芯。使用三点弯曲的生物力学测试来确定每种设计的6个螺钉的极限载荷,屈服强度和破坏循环。 Synthes 4.5毫米空心螺钉的极限载荷,屈服强度和破坏循环显着(P <.05)比Synthes实心空心螺钉低(P <.05),原型空心螺钉中的极限载荷明显低于原型实心螺钉( P <.05)和带有实心芯插入件的原型空心螺钉(P <.05)。带实心芯插入物的原型空心螺钉与实心实心螺钉之间的极限载荷,屈服强度或破坏循环之间无显着差异(P> .05)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号