首页> 外文期刊>Point of care >Performance Evaluation of a Glucose Monitoring System for Point-of-Care Testing with the Critically Ill Patient Population - A Multicenter Study
【24h】

Performance Evaluation of a Glucose Monitoring System for Point-of-Care Testing with the Critically Ill Patient Population - A Multicenter Study

机译:一项针对重症患者人群进行现场护理的葡萄糖监测系统的性能评估-多中心研究

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

To evaluate the performance of the ACCU-CHEK Inform II glucose meter system for use in critically ill patients, and to assess the effects of blood oxygen tension, hematocrit, sodium levels on glucose measurements. A multicenter study involving the collection and testing of remnant deidentified whole blood samples from the blood gas laboratory at 3 institutions. Glucose meter measurements were compared to the Hitachi/Cobas c501 Isotope Dilution Gas Chromatograph aligned glucose hexokinase reference method, and the hospital's laboratory comparison method. Blood gas, whole blood electrolyte, glucose, and hematocrit test results were documented for each sample. Acceptance criteria described in ISO15197:2013 and Clinical Laboratory Standards Institute (CLSI) POCT12-A3 were used to assess clinical performance. Glucose meter results correlated with both the Hitachi/Cobas c501 hexokinase, y = 0.988x - 0.4, R2 = 0.982, N = 579, hospital reference methods, y = 0.974x + 5.5, R2 = 0.972, N = 590. When compared to the glucose hexokinase method, glucose meter results met the International Organization for Standardization and the CLSI acceptance criteria. Glucose measurements were negatively biased at elevated sodium levels, with an approximate bias of -7 (mg/dL, %) at sodium level of 160 mmol/L. The glucose meter system correlated with both glucose hexokinase and hospital laboratory comparison methods, with performance meeting International Organization for Standardization CLSI acceptance criteria. However, meter results were influenced by high blood sodium concentration.
机译:评估用于重症患者的ACCU-CHEK Inform II血糖仪系统的性能,并评估血氧紧张度,血细胞比容和钠水平对血糖测量的影响。一项多中心研究,涉及从3个机构的血气实验室收集和测试残留的未识别全血样本。血糖仪的测量值与Hitachi / Cobas c501同位素稀释气相色谱仪校准的葡萄糖己糖激酶参考方法进行了比较,并与医院的实验室比较方法进行了比较。记录每个样品的血气,全血电解质,葡萄糖和血细胞比容测试结果。使用ISO15197:2013和临床实验室标准协会(CLSI)POCT12-A3中描述的接受标准来评估临床表现。血糖仪结果与Hitachi / Cobas c501己糖激酶相关,y = 0.988x-0.4,R2 = 0.982,N = 579,医院参考方法,y = 0.974x + 5.5,R2 = 0.972,N = 590。用葡萄糖己糖激酶法测定的血糖仪结果符合国际标准化组织和CLSI的接受标准。葡萄糖测量值在钠水平升高时呈负偏差,在钠水平为160 mmol / L时近似为-7(mg / dL,%)。血糖仪系统与葡萄糖己糖激酶和医院实验室比较方法相关,其性能符合国际标准化组织CLSI接受标准。但是,血糖仪的测量结果受血钠浓度高的影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号