首页> 外文期刊>Running algorithm of simulator of wireless distributed measurement-control system with rule based processing >Use of Matlab Package for External Local Calibration of IR Camera with Microbolometric FPA Detector
【24h】

Use of Matlab Package for External Local Calibration of IR Camera with Microbolometric FPA Detector

机译:使用Matlab软件包通过红外热分析FPA检测器对红外热像仪进行外部本地校准

获取原文
获取原文并翻译 | 示例

摘要

This work presents the method of improving the accuracy of temperature measurement results obtained by means of long-wave infrared camera equipped with focal plane array (FPA) detector. The typical accuracy of infrared cameras specified by their manufacturers amounts to ±2 K or ±2% of the measured value of temperature. Very often a better accuracy of the measurements is required, e.g. during the measurements of human body temperature. To improve the measurement accuracy an external local (in terms of temperature range) correction of obtained results on the basis of comparison measurement results of black body temperature and next with use of MATLAB package has been proposed.
机译:这项工作提出了一种方法,该方法可提高通过配备焦平面阵列(FPA)检测器的长波红外摄像机获得的温度测量结果的准确性。制造商指定的红外热像仪的典型精度为±2 K或温度测量值的±2%。通常,需要更高的测量精度,例如在测量人体温度时。为了提高测量精度,已经提出了在比较黑体温度的测量结果的基础上对获得的结果进行外部局部(在温度范围内)校正的方法,然后提出了使用MATLAB软件包的方法。