首页> 外文期刊>Iranian Journal of Nuclear Medicine >New approach for calibration of pixelated scintillation detectors of intraoperative gamma cameras
【24h】

New approach for calibration of pixelated scintillation detectors of intraoperative gamma cameras

机译:术中伽马相机像素化闪烁探测器标定的新方法

获取原文

摘要

Introduction:SURGEOSIGHT is a dedicated intra-operative hand-held gamma camera designed and manufactured by our team to be used for lymphoscintigraphy. Although conventional gamma cameras are widely used, they lack the capability of the sentinel lymph node (SLN) imaging in the operation room. Like conventional gamma cameras it is necessary to calibrate SURGEOSIGHT for linearity distortion, energy and flood correction. Our main objective in this study is to develop and implement three techniques for linearity, energy and uniformity corrections. Methods: The camera is made up of a pixelated cesium iodide (sodium-activated) (CsI(Na)) scintillation crystal with 1 × 1 mm2 element size along with a Hamamatsu H8500 flat-panel multi-anode (64 channels) photomultiplier tube. All triple corrections required lookup tables (LUTs). To generate all three LUTs, one uncalibrated acquisition is necessary. Linearity distortion correction was based on localization of physical pixels of crystal and thin plate spline interpolation to estimate the amount of distortion. Energy correction LUT was simply a photo-peak map of incoming events which was estimated using self-organizing map technique. Results: The results showed improvements qualitatively and quantitatively. The extrinsic energy resolution was enhanced from 50% to about 20.6%. Integral uniformity and the differential uniformity, after uniformity correction, in useful field-of-view (UFOV) was measured 9.5% and 4.5%, respectively. Conclusion: Results substantiated that the correction techniques guarantee uniform and accurate output of the SURGEOSIGHT which is desirable for intra-operative localization of the sentinel lymph nodes in breast cancer and also for imaging of other superficial tumors.
机译:简介:SURGEOSIGHT是由我们的团队设计和制造的专用于术中手持式伽玛照相机,用于淋巴闪烁成像。尽管常规的伽玛相机已被广泛使用,但它们缺乏手术室中前哨淋巴结(SLN)成像的功能。像传统的伽马相机一样,有必要校准SURGEOSIGHT以获得线性失真,能量和洪水校正。我们这项研究的主要目标是开发和实施三种用于线性度,能量和均匀度校正的技术。方法:相机由像素尺寸为1×1 mm2的像素化碘化铯(钠激活)(CsI(Na))闪烁晶体以及Hamamatsu H8500平板多阳极(64通道)光电倍增管组成。所有三重校正都需要查找表(LUT)。要生成所有三个LUT,必须进行一次未经校准的采集。线性失真校正是基于晶体物理像素的定位和薄板样条插值来估计失真量。能量校正LUT只是传入事件的峰图,它是使用自组织映射技术估算的。结果:结果表明,从质量和数量上都有改进。外在能量分辨率从50%提高到大约20.6%。在均匀度校正后,在有用视场(UFOV)中的整体均匀度和微分均匀度分别为9.5%和4.5%。结论:结果证实,校正技术保证了SURGEOSIGHT的均匀和准确输出,这对于乳腺癌术中前哨淋巴结的术中定位以及其他浅表肿瘤的成像是理想的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号