首页> 外文期刊>Nuclear Instruments & Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment >Two-dimensional position resolution and correction of position dependence of pulse heights of superconducting series-junction detectors
【24h】

Two-dimensional position resolution and correction of position dependence of pulse heights of superconducting series-junction detectors

机译:超导串联结检测器的二维位置分辨率和脉冲高度的位置依赖性校正

获取原文
获取原文并翻译 | 示例

摘要

One of the drawbacks of the series-junction detector is that signal heights depend on position of incident photon. A series-junction detector with two-dimensional position resolution has been developed. The effective area is 1.5 x 1.5mm~2. The detector was irradiated by α particles and a position resolution of about 50 μm was obtained. Making use of this position resolution, the position dependence of pulse heights can be corrected, such that the energy resolution was improved from 11% to 0.79%.
机译:串联结检测器的缺点之一是信号高度取决于入射光子的位置。已经开发出具有二维位置分辨率的串联结检测器。有效面积为1.5 x 1.5mm〜2。用α粒子照射检测器,获得约50μm的位置分辨率。利用该位置分辨率,可以校正脉冲高度的位置依赖性,使得能量分辨率从11%提高到0.79%。