Energy non-linearity effects in argon gaseous detectors in the region of the Ar K-absorption edge: experimental results

机译:氩气吸收边缘区域中氩气检测器中的能量非线性效应:实验结果

获取原文
获取原文并翻译 | 示例

摘要

The response of an argon proportional counter to X-rays with energy in the range of 1.5-8 keV was investigated. The response to full-energy absorption events was found to follow different trends for X-ray energies below and above the argon K-absorption edge presenting a discontinuity at this energy. An abrupt decrease of (2.2 +- 0.3)% in the detector amplitude at the K-edge (3.203keV) was measured, corresponding to a region within +- (70 +- 10)eV around the edge where the X-ray energy cannot be obtained unambiguously. This discontinuity corresponds to an abrupt increase of about 0.5 eV in the argon w-value and results" in a sudden increase in the energy resolution from about 17% to about 19%. The detector response to escape peak events follows a linear trend over the entire energy range, very close to the response for full-energy absorption of X-rays with energies above the K-edge.
机译:研究了氩气比例计数器对X射线在1.5-8 keV范围内的能量的响应。发现对全能量吸收事件的响应遵循在氩K吸收边缘以下和之上的X射线能量的不同趋势,在该能量处呈现不连续性。测量到在K边缘(3.203keV)处检测器振幅突然下降(2.2±-0.3)%,对应于X射线能量边缘附近+-(70 +-10)eV内的区域无法明确获得。这种不连续性对应于氩w值突然增加大约0.5 eV,并导致“能量分辨率突然增加,从大约17%增加到大约19%。检测器对逃逸峰事件的响应遵循线性变化趋势。整个能量范围,非常接近能量在K边缘以上的X射线全能量吸收的响应。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号