【24h】

Development of silicon pad detectors and readout electronics for a Compton camera

机译:开发用于康普顿相机的硅垫检测器和读出电子设备

获取原文
获取原文并翻译 | 示例

摘要

Applications in nuclear medicine and bio-medical engineering may profit using a Compton camera for imaging distributions of radio-isotope labelled tracers in organs and tissues. These applications require detection of photons using thick position-sensitive silicon sensors with the highest possible energy and good spatial resolution. In this paper, research and development on silicon pad sensors and associated readout electronics for a Compton camera are presented. First results with low-noise, self-triggering VATAGP ASIC's are reported. The measured energy resolution was 1.1 keV FWHM at room temperature for the Am-241 photo-peak at 59.5 keV. (C) 2002 Elsevier Science B.V. All rights reserved. [References: 10]
机译:使用康普顿相机对器官和组织中放射性同位素标记的示踪剂的分布进行成像,可以在核医学和生物医学工程中获得应用。这些应用要求使用具有尽可能高的能量和良好的空间分辨率的厚的位置敏感型硅传感器来检测光子。本文介绍了针对康普顿相机的硅垫传感器和相关的读出电子设备的研究与开发。报告了低噪声,自触发VATAGP ASIC的初步结果。对于59.5 keV的Am-241光电峰,在室温下测得的能量分辨率为1.1 keV FWHM。 (C)2002 Elsevier Science B.V.保留所有权利。 [参考:10]