首页> 外文期刊>Mechatronics, IEEE/ASME Transactions on >Initial Evaluation of a Tactile/Kinesthetic Force Feedback System for Minimally Invasive Tumor Localization
【24h】

Initial Evaluation of a Tactile/Kinesthetic Force Feedback System for Minimally Invasive Tumor Localization

机译:用于微创肿瘤定位的触觉/运动力反馈系统的初步评估

获取原文
获取原文并翻译 | 示例
           

摘要

Minimally invasive surgery (MIS), while beneficial to patients, leads to new challenges for surgeons and prevents tumors from being localized using finger palpation. A tactile-sensing system (TSS), consisting of a hand-held tactile-sensing instrument (TSI) with a visualization interface, was developed to assist in intraoperative tumor localization. This paper presents the calibration of the TSI and its integration with a visualization interface that allows palpation forces to be displayed. Experiments were conducted to determine whether providing visual force feedback (VFF) to the user would significantly benefit TSS performance when attempting to locate 10 mm hemispherical tumors in ex vivo bovine liver. The TSS with VFF realized a 33% and 21% relative reduction in average and maximum applied forces, respectively, and a 53% relative increase in detection accuracy when compared to the use of the TSS without VFF. Thus, VFF improves the performance of the TSS and has the potential to help surgeons identify tumors intraoperatively during MIS.
机译:微创手术(MIS)虽然对患者有益,但给外科医生带来了新的挑战,并通过手指触诊防止了肿瘤的局限性。开发了由具有可视化界面的手持式触觉仪(TSI)组成的触觉系统(TSS),以协助术中肿瘤定位。本文介绍了TSI的校准及其与可视化界面的集成,该界面可显示触诊力。进行实验以确定当尝试在离体牛肝中定位10 mm半球形肿瘤时,向用户提供视觉力反馈(VFF)是否会显着改善TSS性能。与不带VFF的TSS相比,带VFF的TSS的平均和最大施加力分别降低了33%和21%,检测精度相对提高了53%。因此,VFF可以改善TSS的性能,并有可能帮助外科医生在MIS期间术中识别肿瘤。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号