Objective: To study the biomechanical feature of a newly designed cervical vertebra internal fixation device and its clinical applications. Methods: Some functional spinal units were fixed respectively with titanium plate, fusion cage and new device designed by ourselves, then a controlled biomechanical study including flexion, extension, torsion and lateral bending was performed and the results were analyzed. Results: As to the mechanical performance, fusion cage showed poor performance in extension test and so did the titanium plate in the distortion test. However, the new device showed good performance in every test. Conclusion: Both simple titanium plate fixation and simple fusion cage fixation have biomechanical defaults, but they are complementary. The titanium plate-interbody fusion cage avoids the defaults and has specific advantages.
机译:目的:研究新型颈椎内固定器的生物力学特性及其临床应用。方法:将部分功能性脊柱单元分别用钛板,融合器和自行设计的新装置固定,然后进行包括弯曲,伸展,扭转和侧弯的受控生物力学研究,并对结果进行分析。结果:就机械性能而言,融合笼在拉伸试验中表现较差,而钛板在变形试验中也较差。但是,新设备在每次测试中均显示出良好的性能。结论:简单的钛板固定和简单的融合笼固定均具有生物力学默认值,但它们是互补的。钛板-椎间融合器避免了默认情况,并具有特殊的优势。