Objective: To study the biomechanical feature of a newly designed cervical vertebra internal fixation device and its clinical applications Methods: Some functional spinal units were fixed respectively with titanium plate, fusion cage and new device designed by ourselves, then a controlled biomechanical study including flexion, extension, torsion and lateral bending was performed and the results were analyzed. Results: As to the mechanical performance, fusion cage showed poor performance in extension test and so did the titanium plate in the distortion test. However, the new device showed good performance in every test. Conclusion: Both simple titanium plate fixation and simple fusion cage fixation have biomechanical defaults, but they are complementary. The titanium plate-interbody fusion cage avoids the defaults and has specific advantages.
机译:目的:研究新设计的颈椎内固定装置的生物力学特征及其临床应用方法:一些功能性脊柱单元分别用钛板,融合笼和自称设计的新装置固定,然后是一个受控生物力学研究,包括屈曲,进行延伸,扭转和横向弯曲,分析结果。结果:至于机械性能,融合笼表明延伸试验中的性能差,钛板在变形试验中也是如此。但是,新设备在每次测试中都显示出良好的性能。结论:简单的钛板固定和简单的融合笼固定都有生物力学违约,但它们是互补的。钛板间融合笼避免了默认值并具有特定的优势。