首页> 外文期刊>Journal of dairy science >Modified versus producer milk calibration: mid-infrared analyzer performance validation.
【24h】

Modified versus producer milk calibration: mid-infrared analyzer performance validation.

机译:修改过的牛奶与生产商的牛奶校准:中红外分析仪性能验证。

获取原文
获取原文并翻译 | 示例
       

摘要

Our objective was to determine the validation performance of mid-infrared (MIR) milk analyzers, using the traditional fixed-filter approach, when the instruments were calibrated with producer milk calibration samples vs. modified milk calibration samples. Ten MIR analyzers were calibrated using producer milk calibration sample sets, and 9 MIR milk analyzers were calibrated using modified milk sample sets. Three sets of 12 validation milk samples with all-laboratory mean chemistry reference values were tested during a 3-mo period. Calibration of MIR milk analyzers using modified milk increased the accuracy (i.e., better agreement with chemistry) and improved agreement between laboratories on validation milk samples compared with MIR analyzers calibrated with producer milk samples. Calibration of MIR analyzers using modified milk samples reduced overall mean Euclidian distance for all components for all 3 validation sets by at least 24% compared with MIR analyzers calibrated with producer milk sets. Calibration with modified milk sets reduced the average Euclidian distance from all-laboratory mean reference chemistry on validation samples by 40, 25, 36, and 27%, respectively for fat, anhydrous lactose, true protein, and total solids. Between-laboratory agreement was evaluated using reproducibility standard deviation (sR). The number of single Grubbs statistical outliers in the validation data was much higher (53 vs. 7) for the instruments calibrated with producer milk than for instruments calibrated with modified milk sets. The sR for instruments calibrated with producer milks (with statistical outliers removed) was similar to data collected in recent proficiency studies, whereas the sR for instruments calibrated with modified milks was lower than those calibrated with producer milks by 46, 52, 61, and 55%, respectively for fat, anhydrous lactose, true protein, and total solids..
机译:我们的目标是在使用生产者牛奶校准样品与改良牛奶校准样品进行校准时,使用传统的固定过滤器方法确定中红外(MIR)牛奶分析仪的验证性能。使用生产商牛奶校准样品组校准了10台MIR分析仪,并使用改良的牛奶样品组校准了9台MIR牛奶分析仪。在3个月内测试了三组具有全实验室平均化学参考值的12个验证乳样品。与使用生产者牛奶样品进行校准的MIR分析仪相比,使用改良牛奶对MIR牛奶分析仪进行的校准提高了准确性(即与化学成分更好地吻合),并改善了实验室之间对验证牛奶样品的一致性。与用生产商牛奶套装校准的MIR分析仪相比,使用改良牛奶样品对MIR分析仪进行的校准将所有3个验证集的所有组件的总体平均欧氏距离降低了至少24%。改良牛奶组的校准分别使脂肪样品,无水乳糖,纯蛋白质和总固体样品与验证样品上所有实验室平均参考化学物质的平均欧氏距离分别降低了40%,25%,36%和27%。使用可再现性标准偏差(sR)评估实验室之间的一致性。验证数据中单个Grubbs统计离群值的数量对于用生产者牛奶校准的仪器要比对改良牛奶套装进行校准的仪器高得多(53比7)。用生产者乳校准的仪器的sR(去除了统计异常值)与最近的熟练度研究中收集的数据相似,而使用改良乳校准的仪器的sR低于使用生产者乳校准的仪器的sR 46、52、61和55分别为脂肪,无水乳糖,纯蛋白质和总固体含量的%。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号