首页> 外文期刊>Internal and Emergency Medicine >Does a non-invasive hemoglobin monitor correlate with a venous blood sample in the acutely ill?
【24h】

Does a non-invasive hemoglobin monitor correlate with a venous blood sample in the acutely ill?

机译:非侵入性血红蛋白监测仪是否与急症患者的静脉血样相关?

获取原文
获取原文并翻译 | 示例

摘要

Non-invasive hemoglobin measuring technology has potential for rapid, portable, and accurate way of providing identification of blood loss or anemia. Our objective is to determine if this technology is reliable in critically ill patients presenting to the Emergency Department. Prospective cross-sectional observational study was done at an urban level-one trauma center, 135 subjects were conveniently sampled, suspected of having active bleeding, sepsis, or other critically ill condition. Non-invasive measurements with Masimo (Irvine, CA, USA) Radical-7 and Rad-57 hemoglobin monitors were compared with the Beckman-Coulter LH-550 (Brea, CA, USA) clinical laboratory blood cell analyzer. The primary outcome was the relationship of the non-invasive device to the clinical laboratory results. Secondary evaluations included the effect of pulse rate, systolic BP, respiratory rate, temperature, capillary refill, skin color, nail condition, extremity movement. The Radical-7 was able to capture reading in 78 % (88/113) of subjects, and the Rad-57 in 65 % (71/110) of subjects. The correlation (R2) of the device Hb was 0.69 and 0.67 (p < 00.01) for the Radical-7 and Rad-57, respectively. The coefficient of variation for the Radical-7 was 18 %, and for the Rad57 it was 13 %. Univariate analysis shows none of the observed factors is associated with the difference values between the device Hb and laboratory Hb. Our results show that Radical-7 and Rad-57 devices do not report readings in 29 % of patients and accuracy is significantly lower than reported by the manufacturer with over 50 % of readings falling outside of ± 1 g/dL. We determined that none of the several potential factors examined are associated with the degree of device accuracy.
机译:非侵入性血红蛋白测量技术具有快速,便携式和准确的方法来鉴定失血或贫血的潜力。我们的目标是确定该技术在急诊就诊的危重病人中是否可靠。在城市一级创伤中心进行了前瞻性横断面观察研究,对135名受试者进行了抽样调查,他们怀疑患有活动性出血,败血症或其他重病。使用Masimo(美国加利福尼亚州尔湾)的Radical-7和Rad-57血红蛋白监测仪与Beckman-Coulter LH-550(美国加利福尼亚州布雷亚)临床实验室血细胞分析仪进行了比较。主要结果是无创器械与临床实验室结果之间的关系。次要评估包括脉搏率,收缩压,呼吸频率,温度,毛细血管充盈,皮肤颜色,指甲状况,四肢运动的影响。 Radical-7能够捕获78%(88/113)的受试者的读数,Rad-57能够捕获65%(71/110)的受试者的读数。对于Radical-7和Rad-57,设备Hb的相关性(R2)分别为0.69和0.67(p <00.01)。 Radical-7的变异系数为18%,Rad57的变异系数为13%。单变量分析表明,没有观察到的因素与设备Hb和实验室Hb之间的差异值相关。我们的结果表明Radical-7和Rad-57设备在29%的患者中未报告读数,并且准确性明显低于制造商的报告,超过50%的读数低于±1 g / dL。我们确定所检查的几个潜在因素均与设备准确性的程度无关。

著录项

  • 来源
    《Internal and Emergency Medicine》 |2015年第1期|55-61|共7页
  • 作者单位

    Department of Emergency Medicine San Antonio Military Medical Center">(1);

    San Antonio Uniformed Services Health Education Consortium">(4);

    Tactical Combat Casualty Care Research Program US Army Institute of Surgical Research">(2);

    Department of Emergency Medicine University of Texas Health Science Center at San Antonio">(3);

    Department of Emergency Medicine San Antonio Military Medical Center">(1);

    Department of Emergency Medicine University of Texas Health Science Center at San Antonio">(3);

    Department of Emergency Medicine San Antonio Military Medical Center">(1);

    Department of Emergency Medicine San Antonio Military Medical Center">(1);

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    Point-of-care; Hemoglobin; Analysis; Emergency care; Human;

    机译:护理点;血红蛋白分析;紧急护理;人的;

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号