首页> 外文期刊>IEEE Transactions on Instrumentation and Measurement >NRC-NIST intercomparison of calibration systems for current transducers with a voltage output at power frequencies
【24h】

NRC-NIST intercomparison of calibration systems for current transducers with a voltage output at power frequencies

机译:NRC-NIST用于电流传感器的校准系统的比对,在功率频率下输出电压

获取原文
获取原文并翻译 | 示例

摘要

A number of international comparisons of current transformer calibrations at power frequencies were conducted in the past. However, all comparisons were done with a secondary current output of 5 A. This inter-comparison of current transformer calibration systems for power frequencies between the National Research Council of Canada and the National Institute of Standards and Technology, USA, was done with a primary current range of 100 A and a nonstandard secondary output signal of voltage range of 10 V. The comparison was implemented with a transfer standard consisting of a commercial current-comparator-based transimpedance amplifier, based on a development at the National Research Council of Canada modified to make the output voltage adjustable in magnitude and phase within a range of /spl plusmn/1 percent. The results indicate that there are no significant differences in the overall accuracy of calibration systems for current transducers with a voltage output at power frequencies in each laboratory.
机译:过去,在电源频率下对电流互感器校准进行了许多国际比较。但是,所有比较都是通过5 A的次级电流输出完成的。加拿大国家研究委员会和美国国家标准与技术研究院之间的功率频率电流互感器校准系统之间的比较是通过一次进行的。电流范围为100 A,非标准次级输出信号的电压范围为10V。根据加拿大国家研究委员会(National Research Council)修改后的开发成果,采用基于商用电流比较器的跨阻抗放大器组成的传输标准进行了比较使输出电压的幅度和相位在/ spl plusmn / 1%范围内可调。结果表明,在每个实验室中,以功率频率输出电压的电流传感器的校准系统的整体精度没有显着差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号