This paper describes how to calibrate a four-terminal-pair (4TP) $LCR$ meter at the right (90$^{circ}$) angle. The $LCR$ meter uses the right angle signal to process a synchronous detector as a vector signal receiver. The right-angle calibration method uses, at a minimum, one variable-phase signal generator (SG) and a phase shift circuit. The SG does not require a precision right angle, but it does need good phase stability. The described phase shifter provides the required differential phase of nearly the right angle. As a result, the right-angle phase deviation of the $LCR$ meter is better than 65 $muhbox{rad}$ (0.223 $^{prime}$) with 37-$mu hbox{rad}$ (0.127$^{prime}$) uncertainty at 10 MHz. This method is also applicable to the phase calibration of a vector network analyzer.
机译:本文介绍了如何在直角(90 $ ^ {circ} $)角度校准四端子对(4TP)$ LCR $仪表。 $ LCR $仪表使用直角信号将同步检测器作为矢量信号接收器进行处理。直角校准方法至少使用一个可变相位信号发生器(SG)和一个移相电路。 SG不需要精确的直角,但确实需要良好的相位稳定性。所描述的移相器提供所需的几乎直角的差分相位。结果,$ LCR $计的直角相位偏移优于65 $ muhbox {rad} $(0.223 $ ^ {prime} $)和37- $ mu hbox {rad} $(0.127 $ ^ {在10 MHz时不确定)。该方法也适用于矢量网络分析仪的相位校准。